3-12
Solar 8000M/i patient monitor
2026265-075C
Installation:
Ethernet communication
Ethernet communication
About ethernet
The GE Unity Network uses Ethernet for device-to-device
communications. This local area network links all patient monitors,
clinical information centers, and other GE equipment throughout the
hospital. Depending on the construction of the hospital, thick-net, thin-
net, or CAT-5 twisted pair cabling is used. The Solar 8000M/i is designed
to be used with twisted-pair cabling. Consult GE when trying to interface
with either thick-net or thin-net cabling. The real-time GE Unity
Network operates at 10 Mbps, half-duplex.
Twisted pair
Twisted pair is the most popular cabling because it is easy to install and
flexible to work with. It uses the star topology with a switch as the hub of
the segment. A maximum of 100 meters or 328 feet is the longest length
of twisted pair cable allowed. The maximum number of devices on the GE
Unity Network is 1,000.
Segment
Solar 8000M/i
Solar 8000M/i
CIC Pro
Printer
CIC Pro
Switches
1 to n
041A
Содержание Solar 8000M/i
Страница 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Страница 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Страница 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Страница 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Страница 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Страница 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Страница 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Страница 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Страница 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Страница 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Страница 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Страница 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Страница 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Страница 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Страница 180: ...A 28 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Solar 8000M i display ...
Страница 181: ...2026265 075C Solar 8000M i patient monitor 8000i Patient Monitor B 1 B Appendix B Electromagnetic compatibility ...
Страница 191: ......