TROUBLESHOOTING: Introduction
Revision J
Solar 7000/8000/View Patient Monitor
4-5
414993-001
Configured Memory
Configured memory contains the Ethernet address, internet address, bed
name, care unit name, and standard unit defaults for each monitor.
Continue
CONTINUE is a menu selection on the monitor that appears after a
successful software download. It allows the user to continue downloading
other files without resetting the monitor.
Display On/Off
This front panel switch on the monitor only turns the display ON and
OFF. It has no effect on the software update, but it is used as a screen
saver.
Protected Memory
Protected memory contains a patient’s history and any individualized
changes to the unit defaults.
Power Cycle
To POWER CYCLE or reboot the monitor, turn the power OFF at the
rear of the unit, not at the front switch. A POWER CYCLE initializes the
new software and updates the software revision window.
Service Mode/Menu
There are two service menus available. The service menu found in the
BOOT LOADER is used when downloading the boot code and main
processor code. The SERVICE MODE menu is imbedded in the monitor
MAIN menu and is used for various functions like calibration, video
tests, and downloading monitor interface software.
Warm Start
WARM START is a menu selection on the monitor that appears after a
successful software download. A WARM START will activate the
software previously downloaded. The following also activates a warm
start.
•
Hold down the NBP GO/STOP and ZERO ALL keys on the front
panel. (For the SolarView remote display controller, hold down the
two blank keys.)
•
Press and quickly release the Trim Knob control, NBP GO/STOP and
ZERO ALL keys.
Содержание Solar 7000
Страница 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Страница 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Страница 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Страница 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Страница 20: ...EQUIPMENT OVERVIEW 2 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Страница 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 84: ...MAINTENANCE Checkout Procedure 3 22 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Страница 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Страница 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Страница 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 182: ...CONFIGURATION Advanced User Procedures 6 30 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Страница 184: ...UPPER LEVEL ASSEMBLY 7 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 213: ......
Страница 214: ......