Chapter 4: Installation
In slow-flowing wastewater streams, overflow sampling retains a volume of sample until the Analyzer has a
sample demand. On demand, the Analyzer pulls in sample from the reservoir as excess overflows to a gravity
drain. See
.
An exterior sample peristaltic pump may also be used with this method when there is no sample flow. Up to
four (4) samplers can be operated in parallel using this method when the 5-stream option is selected.
1. Attach and secure the Wastewater sampler brackets to the left side of the Analyzer, as follows. (The
screws are taped to the side of each bracket.)
•
Top bracket
— Slide the following bracket over the plate adapter and secure it by tightening a screw
on each end of the plate adapter.
Figure 23: Top Wastewater Sampler Bracket w/Plate Adapter
•
Lower bracket
— Hold the following bracket so that the screw holes are aligned with the available
screw openings on the Analyzer, and secure with the provided screws. (Depending which openings
were used to install the multi-stream valve bracket to the inside of the Analyzer, the available
openings may vary between (
A
) and (
A
) or (
B
) and (
B
).
Figure 24: Lower Wastewater Sampler Bracket
2. If using the installed external sample pump, the pump should be oriented as shown in
Overflow Method — External Sample Pump Position on page 96.
Pre-drilled screw
holes
Pre-drilled screw holes
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...