Chapter 9: Troubleshooting
To perform the Liquid Reactor Flow test
Press the
Liquid Reactor Flow
button. The button appears orange and the Analyzer displays the various Test
states.
•
Draining
— Pump P1-R empties any liquid that may be in the sparger through the reactor bypass in
preparation for the test.
•
Loading
—The syringe fills with dilution water, which is then dispensed into the sparger. Valve V6-R is
closed and valves V3-R, V4-R, and V5-R are open.
•
Dispensing
— Pump P1-R empties the sparger through the reactor tube.
•
Reactor Flow
— It will take a few seconds for the water to exit the Liquid Waste tube located in the back
of the Analyzer.
•
If you do not see water exiting the waste tube while P1-R is operating, or if the liquid level in the sparger
does not go down, the reactor tube may be plugged.
•
Waiting
— The test is complete.
Gas Leak Checks
There are six tests to help locate leaks in the gas delivery system that may exist in the Analyzer. Step-by-step
instructions for each leak test is provided in the following pages. Perform the procedures in the following
sequential order.
Check 1 — Gas Manifold Leak Check (using the
Gas Manifold
button)
Check 2 — Reactor Inlet Valve Leak Check (using the
Reactor Inlet Valve
button)
Check 3 — Reactor Bypass Leak Check (using the
Reactor Bypass
button)
Check 4 — Reactor Leak Check (using the
Reactor
button)
Check 5 — Reactor Outlet Valve Leak Check (using the
Reactor Outlet
Valve
button)
Check 6 — Gas-Liquid Separator, NDIR Filter, NDIR Sample Path and Sparger Leak Check (using the
Reactor Bypass
button)
Before You Begin
Before starting the Gas Leak Checks, ensure that either an optional air pump is installed or the gas inlet to the
Analyzer is connected to a source of compressed gas (10-20 psig). Refer to
for more
information.
Also, make sure to insert the small plug (included in the Accessories Kit) into the end of the coiled tube coming
from the left side of the NDIR. This prevents leakage of vent gas out of the NDIR during testing.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...