Chapter 4: Installation
Figure 30: Determining Required Purge Flow
If the Air Purifier is not installed, then the Analyzer must be attached to a source of dry, oil-free gas that contains
less than 1 ppb CO
2
, and which is regulated from 10 to 20 psig (69 to 138 kPa). Purified air or nitrogen are
suitable gases.
In either case, press a 1/4-in. OD flexible tube (such as polyethylene tubing) into the
CARRIER GAS INLET
port on
the right side of the Analyzer. The other end of the tube should be connected to the gas source.
Installing the Gas Supply for the IP56 Accessory
If the IP56 accessory is installed, the Analyzer employs dry, oil-free gas to purge the electronics area to prevent
dust accumulation on sensitive components. The tube from the gas source used for the carrier and sparge gas
may be teed to also provide gas to the IP56 accessory. After installing the tee in the gas tubing, press a 1/4-in.
OD flexible tube (such as polyethylene tubing) into the inlet of the Gas Filter mounted on the right side of the
Analyzer. Tubing from the outlet of the filter is connected to the
OPTIONAL GAS INLET
port on the right side of
the Analyzer. The maximum gas pressure is 100 psig (689 kPa).
Warning
Secure all gas cylinders to an immovable structure or permanent wall. Wear eye
protection when using compressed gases to avoid possible eye injury. Only use
regulators designed for use with the selected gas. Store compressed gases in
accordance with all safety codes.
Warning
If the Analyzer is used with compressed gases, the pressure of the gas at the gas inlet must not
exceed 20 psig (138 kPa). Use only nitrogen or CO
2
-free air.
Table A
Table B
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...