![GE Senographe DS Скачать руководство пользователя страница 67](http://html.mh-extra.com/html/ge/senographe-ds/senographe-ds_manual_128685067.webp)
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Product history and compatibility
Page no. 67
Chapter 3
Product-historyDS.fm
4-3
IDC and ADS Software Kit
Different IDC and ADS Software kits are available as a FRU. To determine which IDC and ADS Software
kits you need to order, do the following:
1. Go to the Service Desktop home page (see
2. Determine the system level from the Gantry Board versions displayed on the Service Desktop home
page, and by consulting Table 1 on page 63. The system level will be defined by one of the
System
Level (Model Number)
columns in Table 1 on page 63.
3. Order the appropriate ADS or IDC Software Kit, according to the system level as follows:
System Level
IDC SW Kit
ADS SW Kit
IS4-3 (2383168-5 and lower)
4.2.6 : 2407974-4
53.10.1 (V3 CS) : 5261383-3
43.10.1 (V1/V2 CS) : 5261381-3
Orion1
Nephtys M412s1
Nephtys M413
(2383168-6)
4.2.6 : 2407974-4
53.10.10 : 5261381-4
Penduick (2383168-7)
10.0.5 : 5324092 *
53.40 : 5261381-6
Z400 (2383168-8)
10.0.5 : 5324092
54.10 : 5261381-7
Pixel Spacing
10.0.5 : 5324092
54.11: 5426229
Sunburst Level (2383168-9)
10.0.5 : 5324092
54.20: 5261381-8
* - IDC Software 10.0.5 introduced at Penduick must only be used on the Dione 5 form factor IDC. When
changing an IDC on a pre-Penduick IB system, if no more Dione 4 IDCs exist you will receive a Dione 5 IDC
on which you must install IDC SW 10.0.5.
Note :
If the system has the CESM option, the software versions will be different to those listed in
this table. Refer to the CESM SIP for the CESM version software versions to order.