GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A314 - External CD-RW Unit
Page no. 1439
Chapter 9
JC-DR-A-314.fm
Job Card D/R A314 - External CD-RW Unit
Chapter 9
1
APPLICABILITY
This Job Card is not required if the internal CD-R writer is present.
Applicability code B. See
.
2
SUPPLIES
CD-RW unit
3
TOOLS
Standard Tool Kit
4
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
5
SAFETY PRECAUTIONS
None
6
PREREQUISITES
The installed AWS software must be ADS version 29.0 or later.
7
PROCEDURE
7-1
Configure the CD-RW unit
Switches on the rear panel of the new CD-RW unit must be correctly set before installation. They are not
accessible after the Control Station covers have been installed.
•
Check that the power switch at the rear of the new
CD-RW unit is set to to ON (
I
). A sticker may have
been placed over the switch to prevent it from being
set off.
•
Check that the SCSI ID switch is set to
2
. If neces-
sary, change the setting up or down by inserting a
small flat screwdriver or similar tool into one of the
two slots above and below the switch.