GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Senographe Operations - How to...
Page no. 113
Chapter 3
S2100D How To.fm
Energ
LNRT
PM_yld
ALGORITHM CALIBRATION
Corr
E
PARAM
CALIB
MAGNIFICATION CORRECTION
CALC
OD_mea
CALIB
CALIB REF_ENERGY OF LNRT mAs=(value)
MEASURED DENSITY OD=(value)
NEXT
VALID
CALIB REF. ENERGY calc
VALID
CALC
OD_mea
CALIB
TIME RECIPROCITY FAILURE THICK.=x cm
MEASURED OPTICAL DENSITY=(value)
NEXT
VALID
RECIPROCITY FAIL.:calc
VALID
CHANGE
LNRT PARAM display x=(3 values)
NEXT
LNRT PARAM change x=(value)
VALID
CALC
CALIB
CONFIG
MAGNIFICATION CALIB. Config:contact/magnify
CONTACT CONFIGURATION
YES
NO
CALIB
THICK
MAGNIFICATION CALIB. x cm
Select plexi thickness
2 CM
4 CM
6 CM
MAGNIFICATION CALIB calc
VALID
CHANGE
MAGNIF. PARAM display x=(2 values)
MAGNIF. PARAM change x=(value)
NEXT
VALID
(GENE)
ALGO
LNRT
(TIME RECIP. FAIL.)
MAG.
CORR.
TIME RECIPROCITY FAILURE
CALIB REF_ENERGY OF LNRT
MAGNIFICATION CALIB.
Ill. 6
Ill. 6