Electrotransfer with PWR LARGE SEMIPHOR
TRANSPHOR UNIT and PWR SEMIPHOR TRANSPHOR
UNIT
Introduction
Electrotransfer with PWR LARGE SEMIPHOR TRANSPHOR UNIT and PWR SEMIPHOR
TRANSPHOR UNIT are performed at low current and low voltage, in approximately one
hour.
This section describes how to run the electrotransfer using PWR LARGE SEMIPHOR
TRANSPHOR UNIT and PWR SEMIPHOR TRANSPHOR UNIT.
Recommended current settings
Use the graph to quickly find the maximum current setting for your gel size , or calculate
the gel area (cm
2
) and multiply it by 0.8 mA.
Gel size
(cm
2
)
Maximum
current (mA)
14 × 24
14 × 16
8 × 10
8 × 7
500
400
300
200
100
0
0
100
200
300
400
45
65
180
270
SEMIPHOR TRANSPHOR UNIT series Operating Instructions 29281777 AA
47
5 Operation
5.3 Electrotransfer with PWR LARGE SEMIPHOR TRANSPHOR UNIT and PWR SEMIPHOR TRANSPHOR UNIT