Troubleshooting
2067634-077 Revision B
SEER™ 1000
91
4.
Enter the
Administrator Password
in the second line.
5.
Click
Login
.
The
Hookup
tab opens.
6.
Change to the
Devices
tab.
7.
Select the recorder.
8.
If the recorder to be reset is not listed, click the
M
symbol.
9.
Once the recorder to be reset is listed, select it with a single click
and click
Reset
. A message confirms that the recorder has been
reset.
How to Notify Service if the Recorder is Defective
If the recorder is defective, it will be detected during the self-test when the
recorder is switched on. In case a defect is detected during the self-test,
the following sequence of LED flashes will alert you that a problem has
been detected:
Both LED flash yellow.
Both LED flash green.
A two-second pause follows.
The green
Event Button LED
flashes.
If any of these failures occur, contact GE Technical Support. By counting
the number of flashes, service personnel can determine the cause of the
defect.
Refer to the following table for the defect identified by the number of LED
flashes.