920-504A
© 2013 General Electric Company. All Rights Reserved. Specifications are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product
names mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not affiliated with GE.
g
www.ge-mcs.com
Sensor Size
Temperature Range
T1
NT from -20°C to +120°C (standard)
TA
ET from -45°C to +120°C
T2
ET2 Extended Temperature Range from -45°C to +210°C, maximum operating temperature is +150°C
T3
ET1 Extended Temperature Range from -196°C to +50°C
Pressure Range of Measuring Loops @ 120°C and M1 material
P0
pmax = 66 bar (standard)
PA
pmax = 88 bar
Construction
Type
PF0
Parallel Measuring Loops Seal Less Version
Material of Wetted Parts
M1
Measuring Loops and Connection Part 1.4571 (316Ti) (standard)
62
Measuring Loops and Connection Part 1.4462 (Duplex)
M3
Measuring Loops and Connection Part 2.4602 (Alloy C22)
Process
Connection
D1
Flange DIN DN200/PN16 Form C (EN 1092-1 Form B1)
D2
Flange DIN DN200/PN40 Form C (EN 1092-1 Form B1)
A1
Flange ANSI 8” 150# RF/SF
A2
Flange ANSI 8” 300# RF/SF
A3
Flange ANSI 8” 600# RF/SF
Others on request
RHM 100 - - - -
Basic Order Code RHM 100
Temperature Range
• NT Models from -20 to +120°C (-4 to +248°F)
• ET Models from -45 to +120°C (-49 to +248°F)
• ET2 Models from -45 to +210°C (-49 to +410°F)
• ET1 Models from -196 to +50°C (-320 to +122°F)
(Heating for housing optional, please consult your local
agent)
Electrical Connection
• Junction box aluminum coated (standard). Junction box
in SS 316Ti optional
• Cable entry M25 x 1.5. Optional cable entries M20 x 1.5,
1/2” NPT or 3/4” NPT
• Max cable length between RHM and RHE is 30 m (98 ft.).
Optional 100 m (328 ft.) with special cable
Material of Wetted Parts
• 1.4571 / SS 316 Ti / UNS S31635 (standard)
• 2.4602 / Alloy C22 / UNS N06022
• 1.4462 / Duplex / UNS S31803
• Others on request
Sensor Enclosure/Housing
• Stainless Steel 1.4301 / SS 304, optional in 1.4571 / SS
316Ti. Others on request
• Protection Class IP 65. Optional IP 66 / NEMA 4x
Approvals
• ATEX Ex II 1 G, EEx ia IIC T6-T1
• CSA USA-Canada, Class I, Div. 1, Groups A, B, C, D
• PED according to directive 97/23/EC: Module B + C1
• Others on request
General Specifications RHM 100