Performance RHM 03
Max Flow Rate and Qnom (*) = 5 kg/min (11 lb/min)
Standard Models
Flow Rate
Uncertainty
kg/min
lb/min
in % of reading
5.00
11.02
0.20
3.00
6.61
0.20
1.00
2.20
0.20
0.25
0.55
0.20
0.10
0.22
0.50
Goldline Models (**) - selected sensors
Flow Rate
Uncertainty
kg/min
lb/min
in % of reading
5.00
11.02
0.10
3.00
6.61
0.10
1.00
2.20
0.10
0.50
1.10
0.10
0.25
0.55
0.12
Low Flow Models (**) - selected sensors
Flow Rate
Uncertainty
kg/min
lb/min
in % of reading
2.00
4.41
0.20
1.00
2.20
0.20
0.50
1.10
0.20
0.10
0.22
0.20
0.075
0.17
0.50
Repeatability
Better ± 0.1% of rate, Goldline 0.05%
Temperature
Better ± 1°C
(*) Nominal flow Qnom refers to roughly 8 - 10 m/s (26 - 33
ft/s) velocity in measuring loops for best performance.
(**) Selected sensors are only available in combination with
temperature options T1, TA, standard material and pressure
range.
• Serial/single path versions offer the same accuracy at half
the flow (Qmax serial version = 2.5 kg/min).
• No relevant pressure effect due to torsional oscillation and
semi-circle (non-deforming) measurement section.
• Uncertainty of reading (incl. zero drift) indications refer
to reference conditions H
2
O, 18-24°C (66 - 76°F), 1 - 3
bar (15 – 45 psi) and installation according to field
manual.
• Pressure drop indications refer to H
2
O, with parallel
measuring loops type P1 and standard manifold block
connections.
• For calibration to customer range and / or with
improved uncertainty, please consult factory.
Uncertainty % DP (bar)
Flow Rate kg/min
Uncertainty % DP (bar)
Flow Rate kg/min
Uncertainty % DP (bar)
Flow Rate kg/min
6