Enhance is available in the
Primary Recon
and
Secondary Recon
parameter collections, if the
slice thickness is 0.625 mm or 1.25 mm. The interval for these slice thicknesses must be equal
to the slice thickness (0.625 or 1.25) or one half the slice thickness (0.312 or 0.625).
Illustration 23: Image Enhance images
An "
E
" displays on the image next to the annotation for to indicate IQ Enhance is enabled for the
acquisition.
NOTE:
IQ Enhance is not compatible with Axial, Cine or Cardiac scan types.
6.1.7 Quality Check (QC) Images
QC images are non-diagnostic, preview images generated by the system to check if the scan
acquisition was successful, prior to full IQ images. QC images are displayed in the two
viewports on the scan monitor. QC images have an annotation at the bottom to indicate that
they are not for diagnostic use. QC images are only generated once during a scan series, and
are not saved or available after the scan series is acquired.
QC Images will be auto-generated with the Primary Recon for the following scan types: Axial ,
Cardiac Axial, Helical and Cine.
6.1.8 Flip / Rotate
The reconstructed orientation of an image can be changed from the default mode by selecting
one of the following flip/rotate options in the
Primary Recon
and
Secondary Recon
collections.
•
None
•
FLR (Flip Left Right)
•
FTB (Flip Top Bottom)
•
FLR / FTB (Flip Left Right / Flip Top Bottom)
NOTE:
Flip / Rotate options cannot be selected for images being acquired in the decubitus
patient positions.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
290
6 Recon Parameters
Содержание Revolution CT
Страница 1: ...Revolution CT User Manual OPERATING DOCUMENTATION 5480385 1EN Revision 1 ...
Страница 2: ...This page left intentionally blank Revolution CT User Manual Direction 5480385 1EN Revision 1 2 ...
Страница 216: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 216 2 Patient Schedule ...
Страница 370: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 370 7 SnapShot Freeze ...
Страница 440: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 440 4 Exam Split ...
Страница 471: ...Illustration 18 VR Settings screen Revolution CT User Manual Direction 5480385 1EN Revision 1 Chapter 16 Reformat 471 ...
Страница 502: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 502 7 Batch Film ...
Страница 674: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 674 1 Operator Messages ...
Страница 680: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 680 1 Units of Measure ...
Страница 688: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 688 1 Lexicon ...
Страница 692: ...Illustration 2 GE XXL Revolution CT User Manual Direction 5480385 1EN Revision 1 692 1 Auto Gating Profiles ...
Страница 694: ...Illustration 4 GE Gen CardioVasc Revolution CT User Manual Direction 5480385 1EN Revision 1 694 1 Auto Gating Profiles ...
Страница 696: ...Illustration 6 GE CACS Revolution CT User Manual Direction 5480385 1EN Revision 1 696 1 Auto Gating Profiles ...
Страница 701: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 701 ...