5 Timing Parameters
5.1 Timing Parameters collection
From the
Scan Settings
window, click
Timing
to display the
Timing Parameters
collection.
Illustration 22: Timing Parameters collection
5.2 Timing parameters workflow
Use this procedure to set up the timing parameters for the scan prescription. The patient’s
condition has a significant effect on timing. Make sure that you know how long the patient can
hold her breath. This helps when working with Auto Voice.
It is important to use scan delays for contrast injections. You can only inject once and you must
get it right the first time. Different anatomy enhances at different rates. Check with your
radiologist for the right scan delays for injectors.
1. From the
Scan Settings
window, click
Timing
.
2. If you need a scan delay before the injection starts, set a
Prep Delay
time.
○
The range is 0 to 300 seconds for the first group and 600 seconds for any additional
groups.
○
Auto Voice automatically sets Minimum Prep Delay time based on message length and
the Preset Delay time set for the
Voice, Lights, Timer
selection.
○
Breathing Lights / Timer sets a Minimum Prep Delay time.
3. For an Axial scan type, set an
Inter Scan Delay
time.
4. Set the
Voice, Lights, Timer
options. See
Section 7.3, Set the Voice/Lights/Timer options
for
details.
○
10 languages are available for preset Auto Voice messages 1, 2, or 3. The user can
define additional languages.
○
Preset Delay Time sets a delay from the time the pre-message ends until the X-ray turns
on. This is helpful for patients who need a longer time to follow breathing instructions.
5. To start the scan, click [Confirm Settings].
5.2.1 Set Prep Delay time
Prep Delay time establishes how long the system waits before turning the X-ray on for a given
group of scans. Use this procedure to set a Prep Delay time for giving contrast with a timed
delay.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 11 Scan
287
Содержание Revolution CT
Страница 1: ...Revolution CT User Manual OPERATING DOCUMENTATION 5480385 1EN Revision 1 ...
Страница 2: ...This page left intentionally blank Revolution CT User Manual Direction 5480385 1EN Revision 1 2 ...
Страница 216: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 216 2 Patient Schedule ...
Страница 370: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 370 7 SnapShot Freeze ...
Страница 440: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 440 4 Exam Split ...
Страница 471: ...Illustration 18 VR Settings screen Revolution CT User Manual Direction 5480385 1EN Revision 1 Chapter 16 Reformat 471 ...
Страница 502: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 502 7 Batch Film ...
Страница 674: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 674 1 Operator Messages ...
Страница 680: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 680 1 Units of Measure ...
Страница 688: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 688 1 Lexicon ...
Страница 692: ...Illustration 2 GE XXL Revolution CT User Manual Direction 5480385 1EN Revision 1 692 1 Auto Gating Profiles ...
Страница 694: ...Illustration 4 GE Gen CardioVasc Revolution CT User Manual Direction 5480385 1EN Revision 1 694 1 Auto Gating Profiles ...
Страница 696: ...Illustration 6 GE CACS Revolution CT User Manual Direction 5480385 1EN Revision 1 696 1 Auto Gating Profiles ...
Страница 701: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 701 ...