System settings
Each installed instrument has a set of default parameter values, called system settings.
The System Settings dialog in System Control is used to view and edit the system setting
for the currently selected instrument before the run is started. Follow the instructions
below to change the System Settings.
Action
Step
In the System Control module, select System:Settings.
1
Result: The System Settings dialog opens with the Instructions displayed.
An example is shown below.
Select the instruction to edit from the list. Click the + symbol to show the in-
structions for each category. The instructions in each category differ depend-
ing on the instrument configuration.
2
Select settings and choose parameter values for the selected instruction.
Click OK. Settings will apply until they are changed.
3
To return to the default values defined in the instrument configuration, click
Set Parameters To Strategy Default Values.
4
Manual instructions
It is possible to interact manually with an ongoing run using Manual instructions. Follow
the instructions below to perform manual instructions.
It is also possible to interact with the system manually directly from the Process
Picture.
Note:
70
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD
3 System description
3.6 UNICORN software overview
3.6.3 System control