5-13Compression Device
The compression device is not intended to be used for patient support or restraint.
It may be engaged at any point along the table top and consists of the front
bracket with ratchet, roller and band, and the rear anchor bracket. To use:
1.
Position the patient
2.
Engage the brackets on the front and rear edges of the table top.
3.
Pull up the ratchet handle, pull out the band and hook the rod on the rear
bracket.
4.
Tighten the band by “cranking” the ratchet handle i.e., move it back and forth.
5.
Release the patient by lifting up the ratchet handle and disengaging the band.
ILLUSTRATION 2-5
COMPRESSION DEVICE
5-14Myelographic Operation
For myelographic procedures, the Horizontal Stop selection located on the table
control panel should be off. This inhibits accidental motion of the cone during the
procedure.
To set the myelographic stop:
1.
Press the button, move stop up or down and release it.
2.
Then move the spot film device against it to make sure it is firmly seated.
NOTE:
Engage the compression lock on the IDD as a precaution.
SET THE MANUAL MYELOGRAPHIC STOP ON THE FLUORO
CARRIAGE (
) BEFORE BEGINNING THE
PROCEDURE.