Alarms
About alarm messages
For all alarm messages, refer to the tables listed at the end of this manual. The tables
list the actual alarm message texts, their location on screen, their possible causes, and
suggestions for actions.
Auditory alarm indicators
Alarm signals indicate that an alarm condition is present. The alarm priority levels
are also indicated.
The following table lists alarm signals for different alarm priority levels with the audible
tone pattern (IEC 60601-1-8).
Priority level
High
Medium
Low
Informational
Repeats a pattern of
5-beep tones played two
times
Repeats pattern of
3-beep tones
1-beep tone repeatedly
None
Alarm settings on the Central Viewer
By default, limits for physiological alarms are determined by the profile selected
during patient admission.
Alarm settings for an individual patient can be changed from the default. The
following setting adjustments are available:
●
SpO2 high
,
SpO2 low
●
PR high
,
PR low
●
RR high
,
RR low
●
Apnea delay: seconds
Alarm priorities and alarm delays are not adjustable on the Central Viewer, they can
only be adjusted in the Config UI. The critical low limit for SpO
2
is indirectly adjustable
on the Central Viewer: if you change the
SpO2 low
alarm limit, the critical low limit will
also change.
Adjusting alarm limit settings
To adjust alarm settings for an individual patient:
1.
Open the Single Patient View for the patient.
2.
Open the
Edit Alarm Settings
form:
a. Select the patient’s name to show the actions menu and select
Edit Alarm
Settings
, or
b. Select the high-low limit displayed above the numeric (only available if a
patient is already assigned to a Hub and is undergoing monitoring).
5815148-01
Portrait
™
Mobile Monitoring Solution
85
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...