Cleaning, disinfection, and care
7.
Rinse the device thoroughly using a soft, clean non-linting cloth saturated with
lukewarm tap water (temperature range 27 to 44°C, 81 to 111°F).
8.
Allow the device to air dry until it is visually dry, or dry with thoroughly with a
soft, clean, non-linting cloth.
Drying times may vary based on the environmental conditions. Do not let fluid
pool around connection pins to avoid corrosion. If this happens, blot dry with a
cotton swab or soft cloth.
●
Pay special attention to metal parts (marked here with red) when drying device
to avoid corrosion:
Post-cleaning inspection
Inspect devices visually after cleaning and prior to disinfection for the following:
●
Cleanliness
■
If there is any visible soil present, repeat the cleaning procedure until no visible
soil remains.
●
Damage, including, but not limited to, corrosion (rust, pitting), discoloration,
excessive scratches, flaking, cracks, and wear
●
Missing or worn device labeling or marking
Do not use devices that are not functioning properly or are damaged, or have
unrecognizable markings, or missing or worn device labeling or marking.
Cleaning procedure for the alarming unit
NOTE
If you need to disconnect the alarming unit for cleaning,
remember to reconnect it after cleaning.
Initiate cleaning daily and if any visible soil is present. Remove any soil as soon as
possible by wiping the device to prevent the drying of soil on the device as described
in following procedures.
Soiled devices must be separated from non-contaminated devices to avoid
contamination of personnel or surroundings.
Preparing the alarming unit for cleaning
1.
Visually ensure that the device is intact. Evidence of damage may include, but is
not limited to, discoloration, excessive scratches, wear, and cracks.
5815148-01
Portrait
™
Mobile Monitoring Solution
125
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...