Model PM880 Portable Hygrometer User’s Manual
4.5 Setting a Probe Calibration Reminder
The
Probe ID
command enables you to enter the probe serial number,
calibration dates and a calibration reminder. Use the following steps to enter
probe data:
4.5.1 Accessing the Probe ID Tab
1.
If the menu is not active, press
[MENU]
.
2.
Use the arrow key to scroll to
Program
and press
[ENTER]
.
3.
Use the arrow key to scroll to
Calibrate
and press
[ENTER]
.
4.
Use the arrow keys to scroll to the
Probe ID
tab and press
[ENTER]
. A
screen similar to Figure 20 appears.
Figure 20: Probe ID Window