Chapter 2. Taking Measurements
26
Model PM880 Portable Hygrometer User’s Manual
2.5 Operating the Sample System
Refer to Figure 13 while performing this procedure.
1.
Crack open the sample inlet valve allowing a low volume purge of the
sample system.
2.
Very slowly
open the inlet value allowing it to reach process pressure.
3.
Begin closing the filter/coalescer outlet needle valve and the sample
outlet valve until both valves are
nearly
closed.
4.
Adjust the sample outlet needle valve and the filter/coalescer outlet
valve until you feel a slight pressure against your finger at the venting
tubes.
5.
Open the inlet valve fully. Be sure the flow out of the venting tubes
remains the same as in step 4. To prevent physical damage, do not
subject the probe to excessive flow.
6.
Turn on the PM880.
Figure 13: Sample System Valve Locations
Содержание PM880
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 14: ...Contents xii Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 34: ...Model PM880 Portable Hygrometer User s Manual 20 Chapter 1 Getting Started no content intended for this page ...
Страница 108: ...Chapter 5 Logging Data 94 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 196: ...Appendix B Menu Maps 182 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 202: ...Model PM880 Portable Hygrometer User s Manual 188 Appendix B Menu Maps no content intended for this page ...
Страница 212: ...Index 198 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 217: ......