Chapter 8. Maintenance and Troubleshooting
Model PM880 Portable Hygrometer User’s Manual
161
8.14Recalibrating the Pressure Sensors
Since the pressure sensor on a TF Probe or on Moisture Image Series (MIS)
Probes is a strain gage type, the pressure calibration is linear and is
calibrated at 2 data points. Each point consists of a voltage or FP value and
a corresponding pressure value in psig. Check or change the two calibration
data points using the steps below.
1.
Set one of the boxes on the screen to display pressure in mV (or FP) and
another box to display psig. Refer to
Displaying Measurements
on
page 15 to set up the screen.
2.
Expose the pressure sensor to air and record the mV (or FP) and psig
readings. This reading is for the zero pressure.
3.
Expose the pressure sensor to a known full scale pressure source and
record the mV (or FP) and psig readings. This reading is for the span
pressure.
4.
Press
[MENU]
.
5.
Use the arrow key to scroll to
Program
and press
[ENTER]
.
6.
Use the arrow key to scroll to
Calibration
and press
[ENTER]
.
7.
Proceed to
Entering Pressure Calibration Data
on page 14 to complete
entering data.
Содержание PM880
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 14: ...Contents xii Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 34: ...Model PM880 Portable Hygrometer User s Manual 20 Chapter 1 Getting Started no content intended for this page ...
Страница 108: ...Chapter 5 Logging Data 94 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 196: ...Appendix B Menu Maps 182 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 202: ...Model PM880 Portable Hygrometer User s Manual 188 Appendix B Menu Maps no content intended for this page ...
Страница 212: ...Index 198 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 217: ......