Specifications
10-5
August 2003
Dew/frost Point Temperature
Overall Calibration
Range:
-110°C to 60°C
Calibration Range
Options:
Standard:
20° to -80°C with data to -110°C
Ultralow:
-50 to -110°C
Extended High:
60° to -80°C with data to -110°C
Accuracy:
±2°C from 60° to -65°C
±3°C from -65° to -110°C
Repeatability:
±0.5°C from 60° to -65°C
±1°C from -65° to -110°C
Operating Pressure:
5 µ of Hg to 5,000 psig (limited by
pressure sensor — see pressure
measurement specifications)
Temperature Measurement
Note:
An optional thermistor is available for all GE
Panametrics moisture probes.
Range:
-30°C to 70°C
Accuracy:
±0.5°C at -30°C
Содержание PM880