background image

© 2013 General Electric Company. All Rights Reserved. Specifications are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product 
names mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not affiliated with GE.

www.ge-mcs.com

920-007H

Moisture Measurement

Compatibility

Compatible with all GE aluminum oxide moisture probes:

M Series, TF Series, and Moisture Image Series. Each

probe type requires a different type of I/O cable.

Calibration

GE moisture sensors are computer-calibrated to National 

Institute of Standards and Technology (NIST) traceable 

moisture concentrations.

Dew/Frost Point Temperature

Overall Calibration Range Capability

-166°F to 140°F (–110°C to 60°C)

Calibration Range Options

•  Standard: 68°F to -112°F (20°C to –80°C) with data to 

-166°F (–110°C)

•  Ultralow: -58°F to -166°F (–50°C to –110°C)

•  Extended high: 140°F to -112°F (60°C to –80°C) with 

data to -166°F (–110°C)

Accuracy

•  ±3.6°F (±2°C) from 140°F to -85°F (60°C to –65°C)

•  ±5.4°F (±3°C) from -85°F to -166°F (–65°C to –110°C)

Repeatability

•  ±0.9°F (±0.5°C) from 110°F to -85°F (60°C to –65°C)

•  ±1.8°F (±1.0°C) from -86°F to -166°F (–66°C to –110°C)

Operating Pressure

5 µ of Hg to 5000 psig (345 bar) limited by pressure 

sensor—see pressure measurement specifications

Temperature Measurement

Optional thermistor available for all GE moisture probes

Range

–22°F to 158°F (–30°C to 70°C)

Accuracy

±0.9°F (±0.5°C) at -22°F (–30°C)

Pressure Measurement

Optional pressure sensor available for TF Series and 

Moisture Image Series moisture probes

Ranges

•  30 to 300 psig (3 to 21 bar)

•  50 to 500 psig (4 to 35 bar)

•  100 to 1000 psig (7 to 69 bar)

•  300 to 3000 psig (21 to 207 bar)

•  500 to 5000 psig (35 to 345 bar)

Accuracy

±1% at full scale

Proof Pressure

Three times span of available range up to maximum 

 

7500 psig (518 bar)

Sample System

The SS880A standard sample system comprises an inlet 

needle valve, a built-in coalescing filter and sample cell 

with a bypass needle valve and venting tube, a pressure 

gauge (various ranges), and an outlet needle valve with 

venting tube.

Wetted Parts

316 SS

Operating Pressure

Configurations available for 300, 500, 1000, 2000, or 

 

3000 psig (21, 35, 69, 207, or 345 bar), dependent on 

pressure gauge)

Maximum Pressure Rating

•  Standard: 3000 psig (207 bar)

•  Optional: 5000 psig (345 bar)

Other Options

•  Inlet pressure regulator, 0 to 500 psig (1 to 345 bar) 

outlet

•  Armored flowmeter, 1.3 to 13 SCFH

•  10-ft (3-m), flexible, polytetrafluoroethylene-lined, 

braided–stainless steel hose with 1/4-inch tube 

connections. Not recommended for dew point 

temperatures below –103°F (–75°C).

Содержание PM880

Страница 1: ... and filling Shipboard applications SF6 circuit breakers Furnace gases heat treating General plant facility maintenance Features Hand held portable design Intrinsically safe IP67 rated Large graphic display Internal data logger IrDA communication with PC Stores up to 60 log site files Simple programming via graphic user interface Compatible with all GE moisture probes Lightweight hand held sample ...

Страница 2: ...ilable on site in the factory or on line GE offers extended warranties and service agreements for the PM880 as well as a range of services including NIST traceable calibration field calibration rentals and moisture surveys PM880 Accessories 1 Portable infrared thermal printer and battery charger 2 Zippered soft carrying case 3 Flexible braided stainless steel hose 4 MIS probe electronics module 5 ...

Страница 3: ...splay Printer Terminal Output Infrared communication port Cables Cable type dependent on probe type M Series TF Series or Moisture Image Series LEMO to bayonet connector Cable Length Standard 10 ft 3 m Optional Consult GE for other lengths Hazardous Area Classification Intrinsic safety certification Baseefa 2001 Ltd II 1 G EEx ia IIC T3 20 C Ta 50 C Baseefa02ATEX0191 and CSA C US Class I Division ...

Страница 4: ...Pressure 5 µ of Hg to 5000 psig 345 bar limited by pressure sensor see pressure measurement specifications Temperature Measurement Optional thermistor available for all GE moisture probes Range 22 F to 158 F 30 C to 70 C Accuracy 0 9 F 0 5 C at 22 F 30 C Pressure Measurement Optional pressure sensor available for TF Series and Moisture Image Series moisture probes Ranges 30 to 300 psig 3 to 21 bar...

Отзывы: