5.4
Calibration of volume information
Below is a step by step procedure for the calibration of the volume information transmit-
ted from the Peristaltic Pump P-1 instrument to the fraction collector when connected
to the remote control.
Action
Step
Set up the chromatography system as desired.
1
Connect Peristaltic Pump P-1 into the fraction collector with a communication
cable.
2
Set a flow rate approximately equal to that desired (see
3
For further information refer to the instruction manuals for preferred fraction
collector, e.g. Frac 920.
4
The calibration remains valid whenever the Peristaltic Pump P-1 instrument
and the same fraction collector are connected via the remote outlet. If a differ-
ent i.d. tubing is used, the volume information should be recalibrated.
Note:
The calibration of the volume information is not valid after long term use of
the tube. Calibrate frequently to get high precision.
Note:
48
Peristaltic Pump P-1 Operating Instructions 28962201 AF
5 Operation
5.4 Calibration of volume information