Chapter 15: Preferences
5495975-1EN Rev.9
15-22
© 2013-2017 General Electric Company. All rights reserved.
Retry Failed Print
The setting is to control whether system shall automatically retry the failed print tasks during software
boot.
If it is On, system shall automatically retry the failed print tasks during software boot.
If it is Off, system shall not automatically retry the failed print tasks during software boot.
Copy Exam
Enabling the Copy Exam function allows exams to be copied between the local databases and network
hosts. It also allows exams to be copied to a CD or DVD.
There is no configuration for this function; it is either enabled (ON) or disabled (OFF).
Follow this process to enable or disable Copy Exam:
1. From the Worklist screen, click [UTILITIES].
2. Select
Preferences > Image Management
.
3. Click Copy Exam [ON] to enable the function.
4. Copy Exam [OFF] disables the function.
5. Click [CLOSE].
Auto Tag (Quality Check)
Enabling Auto Tag (or Quality Check) provides an indicator on the Image Viewer screen to mark an image
of acceptable quality. Auto Print and Auto Delete can be configured to act upon images that have the
Quality Check mark.
There is no configuration for this function; it is either enabled (ON) or disabled (OFF).
Note:
This function must be enabled first in order for the RRA feature to become enabled. Refer to
Chapter 13: Advanced Applications-Repeat/Reject Analysis (RRA) (p. 13-8)
for more information.
Disabling this function automatically disables the RRA feature.
Follow this process to enable or disable Auto Tag:
1. From the Worklist screen, click [UTILITIES].
2. Select
Preferences > Image Management
.
3. Click Auto Tag [ON] to enable the function.
Auto Tag [OFF] disables the function. Disabling Auto Tag will automatically disable RRA as well.
4. Click [CLOSE].