Figure 13-25
Image Status on Acquisition screen
The sub-images appear on the Viewer screen as they are acquired and a pasted image appears
approximately 10 seconds later.
19. Continue with
Chapter 13: Advanced Applications-Image Viewer (p. 13-25)
to view, re-paste, or pro-
cess images.
Alternate Auto Image Paste Acquisition Workflows
The following sections provide instructions for the acquisition of specific anatomy.
Leg Paste (at wallstand)
Improper COI determination on Leg anatomy may result in no image below the ankle. The following pro-
cess correctly determines the COI for Leg anatomy and results in a complete image.
Note:
This process is for Auto Image Paste on the Leg anatomy only. COI determination and patient
positioning for the Spine anatomy remains the same.
1. Position the patient as far posterior on the riser as is comfortable (there is no need to have the
patient’s heel touching the Mylar backing of the patient positioner).
2. Determine the COI at a point that is just posterior to the lateral malleolus.
3. Enter the COI value on OTS screen.