2.3 Room Layout Considerations
2.3.1 Radiation Protection
Because X–Ray equipment produces radiation, you may need to take special precautions or
make special site modifications. The General Electric Company does not make
recommendations regarding radiation protection. It is the purchasers’ responsibility to consult a
radiation physicist for advise on radiation protection in x–ray rooms.
2.3.2 Service Access
Allow appropriate space for service access of equipment. Consult component pre–installation
directions for clearance information.
2.3.3 Clinical Access
Make sure that you plan the room with the following clinical access requirements:
•
Provide easy access to the patient table. Stretchers and other mobile hospital equipment
must reach the table quickly.
•
The layout of the table in the room (PIM) shall make a provision so that the clearance
between the maximum table position (head side) on system axis and any object in the room
(e.g.: wall, device) be greater than 50 cm (19.7 in) or 65 cm (25.6 in) if the Header Extender
is used).
•
Provide sufficient space around the patient table for the unimpeded conduct of CPR
(Cardiac Pulmonary Resuscitation). With the table in this position, the table must be capable
of ro/- 45°
•
Clinicians at the patient table must be able to communicate with assistants in the control
area.
•
There must be an unrestricted view of the video monitors and physiological monitoring
equipment from the vascular table.
•
Operators in the control area must have easy access to the control console. However,
position the controls (including handswitches) so that the operator cannot take exposures
while looking around or standing outside the control booth’s lead glass window.
•
Operators in the control area must have easy access to video recorders and injector
programmers, film and video storage cabinets, and service and operating manuals.
•
Consult customer on the number and location of nonelectrical lines (air, oxygen, vacuum,
water, etc.) in the vascular room.
2.3.4 Peripheral Equipment
Consult hospital personnel regarding additional space requirements for the following types of
hospital equipment:
•
Storage cabinets.
•
Sinks.
•
Oxygen stations.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements