resting on top of the table and the blue stabilizing feet under the table. Tighten the
clamping systems with the four- pronged knob. Test to check that head holder is
securely attached to the table. Adjust the height and angle of the head cradle by
loosening the long handled knob. Once cradle is set at desired position tightened the
long handled top. Place the patient’s head in the head cradle. Use the two velcro©
straps to hold the forehead and chin in place. Additional adjustment maybe made to the
position of the head cradle by loosening and tightening the long handled knob if desired.
○
Clear-Vu Arm Support
○
Mattress and mattress slicker
2.3.7 Preference Cabinet locations
NOTE:
This is applicable for all types of cabets (including PDB).
The clear width of the service area in front of the cabinet doors to insure electrical safety shall
be at least 0.9 m for all cabinets except the system control cabinet (1 meter is necessary for
cabinet).
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements
101