January 2006
2-8
Installation
Input Power
Power wiring enters the case through a gland fitting at the lower right
of the unit and connects to a screw terminal block mounted on the
right side of the case. Wiring of this terminal block is shown in
Figure 2-10 below. The voltage and frequency rating and tolerances,
as well as fusing data, are listed on the bottom of the unit.
Figure 2-10: Wall-Mount AC Power Wiring
Output Wiring
The benchtop outputs are connected to removable terminal blocks on
the rear panel. Figure 2-5 on page 2-5 shows the location of the
benchtop terminal blocks and Figure 2-11 below shows the Slot A
connections.
The Wall-Mount
Optica
input/output terminal blocks are located
inside the front door as shown in Figure 2-8 on page 2-7. Cabling is
brought in through the gland on the bottom of the unit and wired to
the terminal blocks shown in Figure 2-11 below.
Note:
Output programming is described in Chapters 4, 5 and 6.
Figure 2-11: Benchtop and Wall-Mount Input/Output Terminal Blocks
Top
Front
L
Case
N
G
of
AL2
NC
OUT A
0-5
PRESSURE
I in
Benchtop Unit
RTN
4-20
OUT B
0-5
V in
+V
Case
of
Front
NO
ALARM 1
ALARM 2
NC
COM
4-20
NC
NO
NO
COM
Slot A
COM
NO
COM
AL1
NC
+V
Wall-Mount Unit
PRES
I in
RTN
V in
Top
OUT B
OUT A
4-20
4-20
0-5
0-5
Содержание Optica
Страница 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Страница 11: ...Chapter 1 ...
Страница 20: ...Chapter 2 ...
Страница 43: ...Chapter 3 ...
Страница 58: ...Chapter 4 ...
Страница 80: ...Chapter 5 ...
Страница 81: ...Programming the 4x40 Optica Introduction 5 1 Programming Technique 5 2 Programmable Functions 5 4 ...
Страница 94: ...Chapter 6 ...
Страница 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Страница 99: ...Chapter 7 ...
Страница 109: ...Appendix A ...
Страница 114: ...Appendix B ...
Страница 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Страница 118: ...Humidity Equations and Conversion Chart B 3 January 2006 Figure B 1 Graphical Humidity Conversion Chart ...
Страница 119: ...Appendix C ...
Страница 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Страница 122: ...Appendix D ...
Страница 123: ...Chilled Mirror Sensors Introduction D 1 Depression D 1 Measurement Range D 2 Comparing Optica Models D 3 ...
Страница 127: ...Appendix E ...
Страница 128: ...Glossary ...
Страница 130: ...Appendix F ...
Страница 133: ...Appendix G ...
Страница 134: ...Communicating with the OPTICA Using Ethernet Direct Communication G 1 Computer Communication G 1 ...