Introduction
2026419-033E
CIC Pro™
1-9
NOTE
Parts and accessories used must meet all local building and safety requirements.
NOTE
Patient environment is any volume in which intentional or unintentional contact
can occur between the patient and parts of the system or between the patient and
other persons touching parts of the system (IEC 60601-1-1).
Equipment symbols
NOTE
Some symbols may not appear on all equipment.
Symbol
Description
ATTENTION: Consult accompanying documents.
TYPE B APPLIED PART: Non-isolated applied part suitable for intentional external and internal application to the
patient excluding direct cardiac application.
[Medical Standard Definition:] Applied part complying with the specified requirements of IEC 60601-1 Medical
Standards to provide protection against electric shock, particularly regarding allowable leakage current.
TYPE BF APPLIED PART: Isolated (floating) applied part suitable for intentional external and internal application to
the patient excluding direct cardiac application. “Paddles” outside the box indicate the applied part is defibrillator proof.
[Medical Standard Definition:] F-type applied part (floating/isolated) complying with the specified requirements of IEC
60601-1 Medical Standards to provide a higher degree of protection against electric shock than that provided by type
B applied parts.
NOTE
The rating of protection against electric shock (indicated by symbol for CF or BF) is achieved only when used
with patient applied parts recommended by GE.
TYPE CF APPLIED PART: Isolated (floating) applied part suitable for intentional external and internal application to
the patient including direct cardiac application. “Paddles” outside the box indicate the applied part is defibrillator proof.
[Medical Standard Definition:] F-type applied part (floating/isolated) complying with the specified requirements of IEC
60601-1 Medical Standards to provide a higher degree of protection against electric shock than that provided by type
BF applied parts.
Writer door button.
Silence Alarm keyboard key.
CE mark.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......