6-24
CIC Pro™
2026419-033E
Configuration
NOTE
All clinical setting values must be received from the GE CAS or from the nursing
director of the care unit. Consult with your CAS or hospital staff about the use of
these functions and if they should be used with your configuration.
Complete the following procedure to configure the
Force Age
setting on the CIC Pro
center:
1.
If you have not already logged onto Webmin,
2.
Click
Configuration > Set Flags
.
NOTE
The
Set Flags Module
screen shows the current set flags settings, which can
be modified.
3.
In the
FORCE AGE
field on the
Set Flags Module
screen, click
Enable
.
4.
Click
Submit
.
5.
Restart the application.
See Perform safe restart of the CIC Pro center on page 8-
6.
Verify the new settings prior to monitoring patients.
Configure multiviewer alarm audio setting, if applicable
This setting is applicable for
mirror
or out of care unit CIC Pro centers. This setting
will affect in-unit audible alarming.
NOTE
All clinical setting values must be received from the GE CAS or from the nursing
director of the care unit. Consult with your CAS or hospital staff about the use of
these functions and if they should be used with your configuration.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......