Chapter 4. Service and Maintenance
Moisture Target Series 6 Hygrometer User’s Manual
83
4.3 Replacing/Recalibrating Moisture Probes
For maximum accuracy, moisture probes should be returned to the factory
for recalibration every 6–12 months, depending on the application. Under
very severe conditions, more frequent calibrations are recommended
However, under very mild conditions, less frequent calibrations are
necessary. Contact a GE applications engineer for your specific
recommended calibration frequency
All new or recalibrated moisture probes must be installed in accordance
with the instructions in Chapter 2,
Installation
.
IMPORTANT:
To maintain good contact at the terminal block and to avoid
damaging the pins on the wiring connector, pull the connector
straight off (not at an angle) the terminal block. Then, make
the cable connections while the connector is off the unit.
Finally, after the wiring is complete, push the connector
straight onto the terminal block (not at an angle).
After the probe has been installed and wired, enter the probe calibration
curve data as described in Chapter 3,
Operation and Programming
. Each
probe is shipped with its own
Calibration Data Sheet
, which includes the
serial number for that probe.
Содержание Moisture Target Series 6
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii Moisture Target Series 6 Hygrometer User s Manual ...
Страница 90: ...Chapter 3 Operation and Programming 80 Moisture Target Series 6 Hygrometer User s Manual ...
Страница 103: ...Appendix B Menu Maps Moisture Target Series 6 Hygrometer User s Manual 93 Appendix B Menu Maps ...
Страница 104: ...Appendix B Menu Maps 94 Moisture Target Series 6 Hygrometer User s Manual Figure 49 Main Menu Map Using M Series Probe ...
Страница 105: ...Appendix B Menu Maps Moisture Target Series 6 Hygrometer User s Manual 95 Figure 50 Main Menu Map Using VeriDri Probe ...
Страница 106: ...Appendix B Menu Maps 96 Moisture Target Series 6 Hygrometer User s Manual no content intended for this page ...
Страница 126: ...Index 116 Moisture Target Series 6 User s Manual ...
Страница 130: ......
Страница 131: ......