Advanced Programming
3-45
March 2008
Entering Auxiliary Input
Curve Data
Note:
S/N is not available in this menu.
Note:
For linear devices only two data points are required; a
maximum of eight data points may be entered for other
devices. Each data point requires a current or voltage value
with a corresponding scale value in the units you want to
output.
.
Repeat the last three steps to enter the remaining data.
Aux Curve Menu 1
Use the arrow keys to move the
brackets to
AUX1
(or
AUX2
),
depending on where the sensor is
connected, and press
[YES].
[S/N] #PTS CURVE
Aux Curve Menu 1
Use the arrow keys to move the
brackets to
#PTS
and press
[YES].
S/N [#PTS] CURVE
Enter Total #PTS 1
Enter the number of data points
and press
[YES].
Then press
2
[NO]
to return to the Curve Menu.
Aux Curve Menu 1
Use the arrow keys to move the
brackets to
CURVE
and press
[YES].
S/N #PTS [CURVE]
# mA/VDC Units 1
If you want to begin entering
data at the data point displayed,
press the right arrow key to
move the bracket to
X1
(mA or
V). Otherwise, enter the desired
data point number. Press
[YES]
and press the right arrow key.
2 +0.000 0.000
# mA/VDC Units 1
Enter the
mA/VDC
value. Press
[YES]
and press the right arrow
key.
2 +1.000 0.000
# mA/VDC Units 1
Enter the units value and press
[YES].
2 +1.000 20.000