March 2008
3-8
Advanced Programming
Setting Up Modbus
Communications
The
MMS-3
can transmit stored data and displayed readings to a
remote ANSI terminal or a personal computer by using an external
RS232-to-RS485 converter. Any
MMS-3
with software version
MMS3.STD.003.J and above has been modified to report
fundamental hygrometry data using
MODBUS
. The meter can transmit
the ten essential measurements using the
MODBUS Read Holding
Register
function.
To program the
MMS-3
:
Press the
[PROG]
key to enter the user program.
Note:
If you have already entered the user program, see Figure 3-3
on page 3-49 to navigate to the Port Menu.
Note:
The
MMS-3
will display the prompt
“
Enter Node ID
” for
selection of a server ("slave") address.
:
Press
[NO]
to exit the
Node ID
menu and again to exit the
FORMAT
menu.
The
FORMAT
and
BAUD
options appear on the same level of the
programming hierarchy, but the communication mode,
MODBUS
,
must be set before selection of its transmission speed.
Enter Passcode: XXXX
Enter the passcode
.
Programming Menu 1
Use the arrow keys to move the
brackets to
PORT
and press
[YES].
_
[PORT] ALARMS
`
Port Menu 1
Use the arrow keys to move the
brackets to
FORMAT
and press
[YES]
.
BAUD [FORMAT]
Format Menu
Use the arrow keys to select
MODBUS
and press
[YES]
.
PRINTER PANALINK [MODBUS]
Enter Node ID 1
Enter a number between
1
and
247
.
1