Chapter 5. Using the Configuration Menu
66
moisture.IQ User’s Manual
5.2 Calibrating the Probes
For newly-purchased moisture.IQ systems, GE has already entered the required calibration data for all
purchased moisture and oxygen sensors. However, you should verify this data before startup. You will also
need to enter new calibration data if you
install a probe from your stock
or
attach a transmitter to the auxiliary
inputs
.
Note:
For hygrometry, you need to enter calibration data only for
M Series
probes. It is not necessary to enter
calibration data for the
Moisture Image Series Probe
unless you send the probe back to GE for calibration
without its electronics module. If this is the case, enter the calibration data for the re-calibrated probe as
described in this section. The moisture.IQ will automatically download the new calibration data into the
Moisture Image Series Probe
electronics module.
5.2.1 Entering Calibration Data Automatically
For moisture and oxygen probes included with the analyzer, GE pre-installs the corresponding calibration data
files. To enter this calibration data, proceed as follows:
1.
On the
Probe Calibration
screen, select the
Channel Number
where a probe requiring calibration is
attached. Then, tap the
FIND
button to display a list of all included probes by serial number.
2.
Scroll the list to find and select the data file matching the probe to be calibrated, and tap the
Save
button. The calibration data for that probe will be automatically entered into the calibration table.
3.
Tap the
Check
button to use the new data.
5.2.2 Entering Calibration Data Manually
Be sure you have the
Calibration Data Sheets
that are supplied with each GE probe. Each
Calibration Data Sheet
consists of a list of data points that must be entered or verified. Each
Calibration Data Sheet
lists the
corresponding probe
serial number
, as well as the preassigned
channel number
.
Calibration Data Sheets
are
usually packed inside the case of the corresponding probe.
To enter your calibration data, refer to the menu map in
and complete the following
steps:
1.
From the
Configuration Menu
, tap the
Probe Calibration
button to open a
Probe Calibration
screen
similar to
. Note that the serial number of the corresponding probe is listed at the
top of the screen for reference.
Содержание moisture.IQ
Страница 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page ...
Страница 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Страница 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Страница 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Страница 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Страница 98: ...Chapter 6 Setting Up Communications 86 moisture IQ User s Manual no content intended for this page ...
Страница 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Страница 138: ...Appendix C Updating the moisture IQ Firmware 126 moisture IQ User s Manual no content intended for this page ...
Страница 180: ... no content intended for this page ...
Страница 181: ......