Work with the Files List
If you select patient reports for transmission from the collapsed
Files
list, the
message displays on the notification area in the lower, right-side of the screen:
<Destination_Name>: <Job_Status>
.
For example, if the destination name is USB, and the job status is
Failed
, the
status displays as follows:
USB: Failed
.
A tick mark displays in the
Sent
column of the
Files
expanded list for patient
reports successfully transmitted to the default destination.
If
Then
The transmission queue has reached its
maximum limit of 1,000 reports, a message
displays in the notification area that the
transmission queue is full and no additional
reports can be added.
Wait for the reports in the queue to transmit
and try again.
One or more patient reports have been
transmitted to the selected destination, a
message displays in the notification asking
you to confirm if you want to re-transmit the
already transmitted reports.
Perform one of the actions below:
• Select
OK
to re-transmit the patient
report.
• Select
Cancel
to cancel the report
transmission.
Patient information is incomplete in one
or more patient reports selected for
transmission (for example, required fields
are blank or contain invalid data), a message
displays in the notification area indicating
that one or patient reports cannot be
transmitted because of incomplete patient
data.
Perform the steps below:
1.
Edit the incomplete patient report to
enter missing patient data.
2.
Retry transmission.
Print a Stored Patient Report
You can print a copy of a stored ECG patient report in any configured report format
for the selected lead set.
1.
From the Acquisition screen, select the
Files
list.
The
Files
collapsed list opens. You can also select the
Expand
icon to open the
Files
expanded list:
2088531-370-2
MAC VU360
™
Resting ECG Analysis System
91