Recording an ECG or Rhythm
Record a Full Disclosure ECG
Before you begin this procedure, make sure:
• You purchase and enable the Full Disclosure option in the
Options Manager
.
• You enable the Full Disclosure option in the
Setting
screen.
1.
Start a new patient test.
2.
To view a Full Disclosure ECG, click the
Full Disclosure
tab on the Acquisition
screen.
• One lead of the Full Disclosure ECG displays. The Full Disclosure shows the
waveform from the left to right side of the screen.
• The Full Disclosure ECG records for a maximum of 5 minutes. The recording
stops after the 5 minutes is complete.
• The Full Disclosure screen displays 10 lines of ECG data and each line is of 30
seconds.
• A notification message displays on the
Acquisition
screen after the Full
Disclosure ECG records for 5 minutes.
• The Full Disclosure screen displays the previous 5 minutes of Full Disclosure
ECG data.
• If you disconnect the acquisition module, the ECG acquisition stops in 10
seconds.
3.
To change the lead, click
Select Lead
.
All the configured leads display. Select the single lead you want to display on the
screen and on the printed FD reports. If you connect the leads to a patient after
being fully disconnected for at least 30 seconds, all the data will be cleared from
the display.
You can record a 15-lead ECG from the Full Disclosure waveform. To do this,
select
Live
screen and change the mode to
Adult 15 Lead
.
The ECG recording restarts and the selected lead is applied to the Full Disclosure
waveform. Clears all the previous recorded data.
4.
To restart the Full Disclosure ECG, click
Restart
. Deletes all the current waveform
data.
The message
unsaved data will be lost
.
Confirm to proceed
displays.
5.
To record a 10 seconds ECG when in the Full Disclosure screen, do the steps that
follow:
68
MAC VU360
™
Resting ECG Analysis System
2088531-370-2