![GE MAC VU360 Скачать руководство пользователя страница 146](http://html2.mh-extra.com/html/ge/mac-vu360/mac-vu360_operators-manual_5075357146.webp)
Patient Preparation
Table 47: Standard 15–Lead Electrode Placement
Item AHA Label
IEC Label
Description
1
V1 red
C1 red
Fourth intercostal space at the right sternal border.
2
V2 yellow.
C2 yellow
Fourth intercostal space at the left sternal border.
3
V3 green.
C3 green
Midway between location 2 and 4.
4
V4 blue
C4 brown
Mid-clavicular line in the fifth intercostal space.
5
V5 orange
C5 black
Anterior axillary line on the same horizontal level as 4.
6
V6 violet
C6 violet
Mid-axillary line on the same horizontal level as 4 and 5.
7
LA black
L yellow
Left deltoid.
8
LL red
F green
Above the left ankle (alternate placement— upper leg as close
to the torso as possible).
9
RL green
N black
Above the right ankle (alternate placement—upper leg as
close to the torso as possible).
10
RA white
R red
Right deltoid.
11
V4R gray
C4R gray
Right anterior chest opposite of 4.
12
V8 gray
C8 gray
Under left mid-scapular line.
13
V9 gray
C9 gray
Left paraspinal border.
Pediatric Electrode Placement
To acquire a pediatric ECG, use the electrode placement shown in the following
diagram.
146
MAC VU360
™
Resting ECG Analysis System
2088531-370-2