Troubleshooting
Error Message
Error Condition
How to troubleshoot:
Cannot perform action while
printing
You have tried to perform
below actions while printing a
patient report:
• Start a test for a new
patient
• Delete the patient report
• Change speed, gain, or filter
• Access User Menu
• Access
Orders
,
Files
, or
Queue
lists
• Power off, lock, log out,
standby or privacy mode
Perform only allowed actions.
Report Transmission Errors
Table 33: Errors Encountered During Patient Report Transmission
Error Message
Error Condition
How to troubleshoot:
Transmission queue is full. Additional
reports cannot be added to the
queue.
The transmission
queue has reached
its maximum limit
of 1,000 reports.
Wait for the reports in the queue to
transmit and try again.
Unable to transmit. Incomplete
patient data.
Unable to transmit one or more
reports. Incomplete patient data.
One or multiple
patient reports
cannot be added
to the transmission
queue because
required fields
in the patient
demographics are
blank or contain
invalid data.
1.
Edit the incomplete patient
report(s) to enter missing
patient data.
2.
Retry transmission.
Report transmission is in progress.
Delete job from queue to edit.
You are trying
to edit a patient
report that is being
transmitted.
Delete the job from the queue to
continue editing the patient report.
Destination unknown
The destination is
not found.
Reconfigure the destination. See
MAC VU360
™
Resting ECG Analysis
System Setup and Configuration
Manual
.
126
MAC VU360
™
Resting ECG Analysis System
2088531-370-2