Troubleshooting
Message
Cause
Solution
Battery not detected
The battery is not detected by
the device.
Perform below steps:
1.
Remove the battery.
2.
Re-insert the battery.
If the battery is still not
detected, replace the battery.
Contact your GE Healthcare
Service Support representative
if the error persists.
Device date/time is incorrect.
Update.
The date and time set on the
device is not correct.
Select
Adjust
to set the correct
date and time. For more
information, see
Configuring
Date and Time
chapter of
MAC
VU360
™
Resting ECG Analysis
System Setup and Configuration
Manual
.
Keyboard failure
The keyboard is not working.
Contact your GE Healthcare
Service Support representative
if the error persists.
Touchscreen failure
The touchscreen is not working. Contact your GE Healthcare
Service Support representative
if the error persists.
Cannot perform action while
acquiring ECG data
You have tried to perform
below actions while acquiring
an ECG or rhythm:
• Start a test for a new
patient
• Enter or edit patient
information (available for
rhythm)
• Change speed, gain, or filter
(available for rhythm)
• Access User Menu
• Access
Orders
,
Files
, or
Queue
list
• Start ECG
• Power off, lock, log out,
standby or privacy mode
Perform only allowed actions.
ECG Acquisition Errors
The table lists messages that you may encounter while acquiring ECG.
122
MAC VU360
™
Resting ECG Analysis System
2088531-370-2