4-32
MAC
TM
600
Revision G
2047426-002
Maintenance:
Check out Procedure
for normal operation.
10. To verify the correctness of programming, go to
on page 3-6 and check if the displayed ARM Software
Version is the latest or not.
BOOT Software Update
1. Select
Setup
.
2. Enter set up password
1111
.
3. Press
Enter
.
4. Press the
Lead
and
F2
keys together.
5. Enter the service password
7763
.
6. Press
Enter
.
7. Select
Update Primary Boot
using the
down arrow
key.
8. Select
Yes
using the
Enter
and
arrow
keys.
A new window opens, askeing you to confirm whether to continue or
not.
9. Select
Yes
using the
arrow
and
Enter
keys.
10. Select
Return
using the
arrow
key and press
Enter.
11. The system displays the following messages:
Programming Primary boot
Primary boot programming Over
12. Press the
esc
key to exit service only Setup.
NOTE
Boot SW update is applicable only if the Boot version has changed
from the previous version in the device and is not already
programmed during ARM software update.
System Log
1. Select
Setup
.
2. Enter the set up password
1111
.
3. Press
Enter
.
4. Press the
Lead
and
F4
keys together.
5. Enter service password
7763
.
6. Insert the blank SD card without write protection enabled.
7. Press the
Enter
key to store the Sys log or the
esc
key to exit the
screen.
NOTE
The SD card needs to be formatted in FAT 16 file system before use
(manufacturer validated SD card is 2GB Transcend and SanDisk). If
Содержание MAC 600 SPY
Страница 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Страница 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Страница 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Страница 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Страница 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Страница 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Страница 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 41: ...Revision G MACTM 600 3 17 2047426 002 Troubleshooting Equipment Problems Paper Moving But Not Printing ...
Страница 42: ...3 18 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Printer Door Open Message ...
Страница 43: ...Revision G MACTM 600 3 19 2047426 002 Troubleshooting Equipment Problems Printing Only Baselines ...
Страница 44: ...3 20 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Battery Error Message ...
Страница 45: ...Revision G MACTM 600 3 21 2047426 002 Troubleshooting Equipment Problems Unit Not Turning ON in Battery Mode ...
Страница 46: ...3 22 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems External Power LED Not Lighting Up ...
Страница 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Страница 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Страница 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Страница 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Страница 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Страница 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Страница 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Страница 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...
Страница 126: ...Index i 4 MACTM 600 Revision G 2047426 002 ...