Battery Test
Battery tests check the current battery status, battery discharge rate and battery
charge rate. Test results are stored in memory and can be printed out. The graphic
displayed shows the
Battery Test
menu. Each test is covered in detail below.
Battery Status
. This test displays and constantly updates current information on
the battery voltage, battery current, percent of charge remaining and battery
temperature.
Battery Voltage
. With a reading of 80% or more for the percent of charge
remaining, the battery voltage should be between 15 and 24 volts. If battery
voltage is below 15 volts, the battery may need to be replaced.
Battery Current
. Disconnect AC power. If the battery current is less than -
0.7 amps, the main CPU may need to be replaced. For example if current is
-0.8 amps, consider replacing CPU. See
Battery Temperature
. A temperature reading over 45° C indicates a failure.
If battery temperature is more than 10° C over the ambient temperature
consider replacing battery.
Ambient Temperature
. Indicates the temperature inside the unit. The
temperature displayed is accurate to 5°C in the range of 0°C to
50°C. Ambient temperatures not within this range cannot be displayed.
19A
20A
Содержание MAC 3500
Страница 1: ...GE Healthcare MACTM 3500 Resting ECG Analysis System Version 9D Service Manual 2021337 036 Revision L ...
Страница 9: ...Revision L MAC 3500 Resting ECG Analysis System 1 1 2021337 036 1 Introduction ...
Страница 10: ...1 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 19: ...Revision L MAC 3500 Resting ECG Analysis System 2 1 2021337 036 2 Equipment Overview ...
Страница 20: ...2 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 25: ...Revision L MAC 3500 Resting ECG Analysis System 3 1 2021337 036 3 Installation ...
Страница 26: ...3 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 40: ...3 16 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Installation General Assembly For your notes ...
Страница 41: ...Revision L MAC 3500 Resting ECG Analysis System 4 1 2021337 036 4 Troubleshooting ...
Страница 42: ...4 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 44: ...4 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Troubleshooting Assembly Descriptions ...
Страница 83: ...Revision L MAC 3500 Resting ECG Analysis System 5 1 2021337 036 5 Maintenance ...
Страница 84: ...5 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 122: ...5 40 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Maintenance Functional Checkout Procedures ...
Страница 123: ...Revision L MAC 3500 Resting ECG Analysis System 6 1 2021337 036 6 Parts Lists ...
Страница 124: ...6 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 152: ...6 30 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Parts Lists Field Replaceable Units FRUs ...
Страница 153: ...Revision L MAC 3500 Resting ECG Analysis System A 1 2021337 036 A Technical Specifications ...
Страница 154: ...A 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 160: ...A 8 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Technical Specifications Barcode Scanner For your notes ...
Страница 161: ...Revision L MAC 3500 Resting ECG Analysis System B 1 2021337 036 B Software Hardware Compatibility ...
Страница 162: ...B 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 167: ...Revision L MAC 3500 Resting ECG Analysis System C 1 2021337 036 C Electromagnetic Compatibility ...
Страница 168: ...C 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 175: ...Revision L MAC 3500 Resting ECG Analysis System Index 1 2021337 036 Index ...
Страница 176: ...Index 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 178: ...Index Index 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 ...
Страница 179: ......