28 2042376-001 Rev G ©2012 General Electric Company. All rights reserved. Lullaby
TM
Warmer
CAUTION:-
Do not autoclave or gas sterilize the skin temperature probe. Do not immerse the
probe in liquid cleaners. These precautions will prevent damage to the probe.
CAUTION:-
Do not allow cleaning fluid to leak into the probe and electrical connectors. This
may damage the equipment.
Step
Description
1
Determine if the patient probe is disposable or reusable:
Reusable probes use a separate, heat-reflecting patch, are gray and have a
round metal disk at the patient end.
Disposable probes come with a smaller heat reflecting patch already
attached, are white and have no metal disk at the patient end.
2
Clean the reusable patient temperature probe by gently wiping with a soft damp cloth
containing a disinfecting agent safe for use on the probe materials. Always be sure to
wipe dry all cleaning agents after cleaning.
NOTE
Disposable skin temperature probes cannot be cleaned and are intended for single
patient use.
3.8
Cleaning Other Components
CAUTION:-
Do not clean the Lullaby
TM
Warmer with organic solvents, scouring compounds,
strong acids, or strong bases. These compounds may damage the components.
CAUTION:-
To minimize the generation of static electricity, do not polish the bedside panels
with a dry cloth.
CAUTION:-
Do not autoclave or gas sterilize any of the plastic parts.
CAUTION:-
The area of the Lullaby
TM
Warmer that comes into contact with iodophor solution
will turn yellow due to staining. Iodine based solutions tend to discolor the surface
of the Lullaby
TM
Warmer.
NOTE
Do not soak any component of the Lullaby
TM
Warmer in cleaning solutions.
Always wipe dry the components to remove traces of the cleaning solutions.
Following these recommendations will extend the life of the components.
NOTE
Do not allow excess cleaning solution to seep between plastic parts (for example:
between the side panels and the lock or the hinges) where it cannot be easily
wiped dry with a cloth.
Содержание Lullaby Warmer
Страница 12: ...This page is left blank intentionally ...
Страница 138: ...112 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer Hood Assembly Continued ...
Страница 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Страница 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Страница 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...