20 2042376-001 Rev G ©2012 General Electric Company. All rights reserved. Lullaby
TM
Warmer
2.2
Controller Checks
WARNING :-
Do not use the Lullaby
TM
Warmer in the presence of flammable anesthetics: an
explosion hazard exists under these conditions.
WARNING :-
Always connect the Lullaby
TM
Warmer directly to a wall outlet. Connecting to a
power strip or another piece of equipment may result in power failure errors.
Illustration
Description
Figure 37:
Power Connection
Connect the power cord to a rated AC wall socket
and lock the caster brakes.
NOTE: Power outlets are region specific.
Figure 38:
Patient Probe Connection
Connect the patient probe to the jack on the
control panel.
NOTE: Insert the probe completely into the jack till
you hear a click.
Figure 39:
Power Standby/ON Switch
Switch on the Power Standby/ON switch located on
the control panel.
Check that:
All the displays and indicators light
Heater level bar graph is 100%
Audible beeps will be heard
Observation lamp illuminates on when the
Observation Light Standby/ON switch is
operated.
NOTE: Do not press any key on the control panel
when switching on the Lullaby
TM
Warmer. This
will turn on the “System failure” alarm and
“key” is displayed on the set temperature
display (Refer to Alarms and Indications).
Power Standby/ON
Switch
Содержание Lullaby Warmer
Страница 12: ...This page is left blank intentionally ...
Страница 138: ...112 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer Hood Assembly Continued ...
Страница 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Страница 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Страница 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...