Technical Publication

LOGIQ V2/LOGIQ V1

Basic Service Manual

Direction Number:

 

5610739-100 English

Rev.10

All Material Copyright © 2015-2017 by General Electric Company Inc. 
All Rights Reserved.

Содержание LOGIQ V2

Страница 1: ...GE Technical Publication LOGIQ V2 LOGIQ V1 Basic Service Manual Direction Number 5610739 100 English Rev 10 All Material Copyright 2015 2017 by General Electric Company Inc All Rights Reserved...

Страница 2: ...tem Copyright 2015 2017 General Electric Company GE Medical Systems a General Electric Company going to market as GE Manufacturer GE Medical Systems China Co Ltd No 19 Changjiang Road WuXi National Hi...

Страница 3: ...Reason for change Rev 1 2015 06 09 Initial Release Rev 2 2015 11 13 New software version general update Rev 3 2015 11 23 Add software wipe information Rev 4 2016 01 05 Update spare part list Rev 5 20...

Страница 4: ...i 2 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Important precautions Translation policy...

Страница 5: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 3 5610739 100 English Rev 10...

Страница 6: ...i 4 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10...

Страница 7: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 5 5610739 100 English Rev 10...

Страница 8: ...i 6 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10...

Страница 9: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 7 5610739 100 English Rev 10...

Страница 10: ...i 8 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10...

Страница 11: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 9 5610739 100 English Rev 10...

Страница 12: ...i 10 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10...

Страница 13: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 11 5610739 100 English Rev 10...

Страница 14: ...i 12 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10...

Страница 15: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 13 5610739 100 English Rev 10...

Страница 16: ...ithin this 14 day period Certified electrical contractor statement For USA Only All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment...

Страница 17: ...pecific information listing the system type manual title part number revision number page number and suggestion details Mail the information to GE Medical Systems China Co Ltd No 19 Changiang Road Wux...

Страница 18: ...ety requirements refer to Chapter 1 in the Service Manual DANGER DANGEROUS VOLTAGES CAPABLE OF CAUSING DEATH ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING WARNI...

Страница 19: ...uplicated in any form in whole or in part without prior written permission of GE GE may revise this publication from time to time without written notice Trademarks All products and their name brands a...

Страница 20: ...i 18 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10...

Страница 21: ...he Proprietary Service Manual 1 4 LOGIQ V2 LOGIQ V1 models covered by this manual 1 5 Important conventions Conventions used in book 1 6 Standard hazard icons 1 8 Product icons Labels locations Safety...

Страница 22: ...ltrasound system and EchoPAC PC in same room 2 18 Networking setup requirements 2 19 Environmental Dangers Patient Environment IEC60601 1 and ANSI AAMI ES60601 1 2 22 Chapter 3 System Setup Overview C...

Страница 23: ...up Contents in this Section 3 42 User s Manual s 3 42 Product Locator Installation Card 3 42 Chapter 4 General Procedures and Functional Checks Overview Purpose of this chapter 4 2 Contents in this ch...

Страница 24: ...t the Cineloop controls 4 38 Back End Processor checks 4 39 Operator Panel Test 4 39 Peripheral checks 4 39 Turn OFF Power to LOGIQ V2 LOGIQ V1 4 39 Power supply test adjustments Power Supply Test Pro...

Страница 25: ...Customer provided prerequisite 7 13 Data Management moving all images 7 13 Backing up the Patient Archive and System Configurations 7 14 Recording important settings and parameters 7 14 Loading the Sy...

Страница 26: ...sks be performed 10 6 Tools required Standard GE tool kit 10 8 GE 2 tool kit 10 10 Special tools supplies and equipment used for maintenance 10 11 System maintenance Preliminary checks 10 12 Functiona...

Страница 27: ...LOGIQ V2 LOGIQ V1 Basic Service Manual i 25 5610739 100 English Rev 10 Inspection Paperwork Ultrasound Inspection Forms 10 33 Electrical Safety Tests Log Index...

Страница 28: ...i 26 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10...

Страница 29: ...00 English Rev 10 Chapter 1 Introduction This chapter describes important issues related to safely servicing the Ultrasound system The service provider must read and understand all the information pre...

Страница 30: ...Overview on page 1 3 Important conventions on page 1 6 Product icons on page 1 10 Labels locations on page 1 11 Safety considerations on page 1 13 Dangerous procedure warnings on page 1 22 Lockout Ta...

Страница 31: ...Legal notes and the Table of Contents TOC An Index has been included after Chapter 10 Table 1 1 Contents in this manual Chapter number Chapter title Description 1 Introduction Contains a content summ...

Страница 32: ...dures for running diagnostic or related routines for the LOGIQ V2 LOGIQ V1 8 Replacement procedures Provides disassembly procedures and reassembly procedures for all changeable FRU 9 Renewal Parts Con...

Страница 33: ...Q V1 models covered by this manual NOTE When not otherwise specified the contents of this manual applies to all LOGIQ V2 LOGIQ V1 models Table 1 2 LOGIQ V2 LOGIQ V1 Model Designations Model Number Des...

Страница 34: ...ment are described next Model designations This manual covers the LOGIQ V2 LOGIQ V1 Ultrasound systems listed in LOGIQ V2 LOGIQ V1 models covered by this manual on page 1 5 Icons Pictures or icons are...

Страница 35: ...e used to provide important information about an item or a procedure NOTE Be sure to read the notes the information contained in a note can often save you time or effort DANGER Danger is used to indic...

Страница 36: ...on as seen throughout this chapter In addition to text several different graphical icons symbols may be used to make you aware of specific types of hazards that could possibly cause harm Even if a sym...

Страница 37: ...others icons make you aware of specific procedures that should be followed Be sure to read the notes the information contained in a note can often save you time or effort Table 1 4 Standard Icons tha...

Страница 38: ...2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Product icons It is important to refer to the current revision of the Ultrasound system s User Manual for a full list of product labels prior...

Страница 39: ...er Manual for a full list of product labels prior to servicing the system LOGIQ V2 LOGIQ V1 labels are provided in English The labels are at the bottom of the system The label content may be different...

Страница 40: ...identification the Unique Device Identification UDI Label The UDI label consists of a series of alpha numeric characters and barcode which uniquely identify the LOGIQ V2 LOGIQ V1 system as a medical d...

Страница 41: ...re in this manual violates safety standards of design manufacture and intended use of the equipment Human Safety Operating personnel must not remove the system covers Servicing should be performed by...

Страница 42: ...ait for at least 20 seconds for capacitors to discharge as there are no test points to verify isolation The amber light on the OP panel ON OFF button will turn off Ultrasound system components may be...

Страница 43: ...rotection Equipment PPE such as gloves safety shoes safety glasses and kneeling pad to reduce the risk of injury WARNING Beware of possible sharp edges on all mechanical parts If sharp edges are encou...

Страница 44: ...ny patient data images system setups to removable media or hardcopy before doing a software upgrade WARNING Ultrasound probes are highly sensitive medical instruments that can easily be damaged by imp...

Страница 45: ...since it may become unstable and tip over CAUTION Use protective glasses during drilling filing smooth surfaces and during all other work where eyes need protection CAUTION Use safety shoes when doin...

Страница 46: ...e connected to an electrical ground The Ultrasound system is equipped with a three conductor AC power cable This must be plugged into an approved electrical outlet with safety ground The power outlet...

Страница 47: ...system Inspect the probe prior to each use for damage or degradation to the housing cable strain relief lens seal connector pins locking mechanism Do not use a damaged or defective probe Never immers...

Страница 48: ...re Do not expose the battery to temperature over 50 C 122 F Keep it away from fire and other heat sources Do not charge the battery near a heat source such as a fire or heater Do not leave the battery...

Страница 49: ...using it If you have any questions about the battery consult GE or your local representative Short term less than one month storage of battery pack Store the battery in a temperature range between 5 C...

Страница 50: ...e removed from an operating LOGIQ V2 LOGIQ V1 some metal surfaces may be warm enough to pose a potential heat hazard if touched even while in shutdown mode WARNING Explosion Warning DO NOT operate the...

Страница 51: ...ut Devices 6 Control all stored and residual energy 7 Verify isolation All potentially hazardous stored or residual energy is relieved WARNING Energy Control and Power Lockout for LOGIQ V2 LOGIQ V1 Wh...

Страница 52: ...e regulated medical waste for transportation purposes and must be transported as a hazardous material NOTE The USER SERVICE staff should dispose of all the waste properly per federal state and local w...

Страница 53: ...FI EMI and can be radiated through space or conducted over interconnecting power of signal cables In addition to electromagnetic energy EMC also includes possible effects from electrical fields magnet...

Страница 54: ...ion point located on the rear of the Ultrasound system near the power connector Follow general guidelines for handling of electrostatic sensitive equipment WARNING Risk of electrical shock Ultrasound...

Страница 55: ...rt resource as listed below Before you call identify the following information and acquire image Alt D to send to the Customer Care team 1 System ID serial number 2 Software version 3 Date and time of...

Страница 56: ...0732 Latin America Service Application Support 1 800 321 7937 1 262 524 5698 Europe OLC EMEA GE Ultraschall Deutschland Gmbh Co KG Beethovenstra e 239 Postfach 11 05 60 D 42655 Solingen Germany OLC E...

Страница 57: ...sentative AUTHORIZED REPRESENTATIVE TELEPHONE FAX NUMBER The location of the CE marking is shown in the Safety chapter of the User manual Authorized EU Representative European registered place of busi...

Страница 58: ...Introduction 1 30 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10...

Страница 59: ...0 English Rev 10 Chapter 2 Site Preparations This chapter provides the information required to plan and prepare for the setup of an Ultrasound system Included are descriptions of the facility and elec...

Страница 60: ...V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Overview Contents in this chapter Overview on page 2 2 General Ultrasound system requirements on page 2 3 Facility needs on page 2 12 Enviro...

Страница 61: ...ronmental requirements If the Ultrasound system is very cold or hot When unpacking the Ultrasound system allow the temperature of the Ultrasound system to stabilize before powering up The following ta...

Страница 62: ...st be turned off Table 2 2 System Environmental Requirements Operational Storage Transport 16hrs Temperature 10 40 C 50 104 F 5 50 C 23 122 F 5 50 C 23 122 F Humidity 30 80 non condensing 10 90 non co...

Страница 63: ...e room NOTE Each person in the room places an additional 300 BTU h demand on the cooling system Lighting Bright light is needed for Ultrasound system installation updates and repairs However operator...

Страница 64: ...Ultrasound outlet Sites with a mains power system without a defined Neutral The dedicated line shall consist of one phase two lines not shared with any other circuit and a full size ground wire from...

Страница 65: ...ation must meet all current local state and national electrical codes Unit power plug If the Ultrasound system arrives without a power plug or with the wrong plug you must contact your GE dealer or th...

Страница 66: ...1000 4 11 Voltage dips short interruptions and voltage variations on mains supply 5 T 95 dip for 0 5 cycle 5 T 95 dip for 0 5 cycle Mains power quality should be that of a typical commercial or hospit...

Страница 67: ...nstallation Possible EMI sources should be identified before the Ultrasound system is installed Electrical and electronic equipment may produce EMI unintentionally as the result of a defect Some of th...

Страница 68: ...nterfere with the ultrasound signals Replace broken Radio Frequency gaskets If more than 20 or a pair of the fingers on an Radio Frequency gasket are broken replace the gasket Do not turn on the Ultra...

Страница 69: ...reparation checks as soon as possible if possible six weeks before delivery to allow enough time to make any changes CAUTION Ensure that the probe face temperature does not exceed the normal operation...

Страница 70: ...ion Purchaser responsibilities on page 2 13 Required facility needs on page 2 14 Desirable features on page 2 15 Minimal floor plan suggestion on page 2 16 Recommended floor plan suggestion on page 2...

Страница 71: ...ducts involved and the accompanying electrical installations are highly sophisticated and special engineering competence is required All electrical work on these products must comply with the requirem...

Страница 72: ...e the Ultrasound system Sites with a mains power system with defined Neutral and Live The dedicated line shall consist of one phase a neutral not shared with any other circuit and a full size ground w...

Страница 73: ...l are within 2 m 6 5 ft of each other with peripheral within 1 m of the unit to connect cables Power outlets for other medical equipment Power outlets for test equipment within 1 m 3 2 ft of Ultrasoun...

Страница 74: ...ches 2 Film Viewer 3 Counter Top Sink with hot and cold water and Supplies Storage 4 Linen Supply 5 Probes Supplies 6 Examination Table 1930 x 610 mm 76 x 24 inches 7 Footswitch 8 Stool 9 Ultrasound s...

Страница 75: ...Sink with hot and cold water 6 Overhead Lights Dimmer Dual Level Lighting bright and dim 7 Emergency Oxygen 8 Suction Line 9 Ultrasound system 10 Dedicated Power Outlet Circuit Breaker protected and e...

Страница 76: ...plan Ultrasound system and EchoPAC PC in same room Figure 2 3 Suggested Room with EchoPAC PC workstation and Ultrasound Scanner 1 EchoPAC PC workstation parts 2 UPS 3 Ethernet network wall outlet 4 3x...

Страница 77: ...Mbit Interface Purpose of the DICOM network function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network Examples of...

Страница 78: ...here are no spaces in any field of the form See Entries must include A host name local port number AE Title IP address and Net Mask for the Ultrasound system The IP addresses for the default gateway a...

Страница 79: ...Facility needs LOGIQ V2 LOGIQ V1 Basic Service Manual 2 21 5610739 100 English Rev 10 DICOM option setup requirements continued Figure 2 4 Worksheet for DICOM Network Information...

Страница 80: ...h extra protective earth will be required in order to meet IEC60601 1 standards for electrical leakage Sub Clause 3 79 and figure A 9 IEC60601 1 2005 and ANSI AAMI ES60601 1 2005 Such an area is an en...

Страница 81: ...nstall LOGIQ V2 LOGIQ V1 system Included is a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim How to prepare the facility and unit of the actual...

Страница 82: ...Overview on page 3 2 Setup reminders on page 3 3 Receiving and unpacking the equipment on page 3 6 Preparing for setup on page 3 13 Completing the setup on page 3 14 System Configuration on page 3 18...

Страница 83: ...les with optional equipment may take slightly longer Setup warnings DANGER WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE I E METER S GROUND SWITCH IS OPEN DON T TOUCH TH...

Страница 84: ...unit to acclimate before you turn it on It requires one hour for each 2 5 C increment it s temperature is below 3 C or above 40 C The following table describes guidelines for reaching operational temp...

Страница 85: ...s should be fully read and understood before operating the LOGIQ V2 LOGIQ V1 and kept near the Ultrasound system for quick reference CAUTION Acoustic Output Hazard Although the ultrasound energy trans...

Страница 86: ...Q V2 LOGIQ V1 on page 3 6 Unpacking the LOGIQ V2 LOGIQ V1 on page 3 8 Warnings for receiving and unpacking Receiving the LOGIQ V2 LOGIQ V1 Overview Improper handling during transportation may harm the...

Страница 87: ...BEFORE delivery is accepted or signed for by a GE representative or hospital receiving agent 2 Report the damage to the carrier Whether noted or concealed damage MUST be reported to the carrier immedi...

Страница 88: ...ply If shipping damage or shortage occurs contact the address shown in Chapter 1 NOTE Please check the LOGIQ V2 LOGIQ V1 console is well assembly after unpacking the system 1 Cut the adhesive tape and...

Страница 89: ...Unpacking the LOGIQ V2 LOGIQ V1 continued 2 Take out console together with 2 interleavers from console package 3 Take out the interleavers beside Accessories Package 4 Take out Accessories Package Fig...

Страница 90: ...up 3 10 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Unpacking the LOGIQ V2 LOGIQ V1 continued 5 Remove 2 interleavers 6 Remove plastic bag Figure 3 3 Removing interleavers and pl...

Страница 91: ...instructions Figure 3 4 Labels on Package Moving into Position In general a single adult can move the LOGIQ V2 LOGIQ V1 Before moving store all loose parts in original accessory box or in back pack Re...

Страница 92: ...GIQ V2 LOGIQ V1 are recyclable The Transportation Box is made of spruce or similar material PHYTOSANITARY CERTIFICATE included in all shipments to The People s Republic of China Lever lockings hinges...

Страница 93: ...system has been damaged please refer to Damage in transportation on page i 14 in the beginning of this manual EMI protection The LOGIQ V2 LOGIQ V1 has been designed to minimize the effects of Electro...

Страница 94: ...Accessories Connector Panel on page 3 16 Connecting probes on page 3 17 Powering the system on page 3 17 System specifications System requirements verification Verify that the site meets the requirem...

Страница 95: ...duct labels prior to serving the system Electrical specifications for LOGIQ V2 LOGIQ V1 In the table below the electrical specifications for LOGIQ V2 LOGIQ V1 includes monitor and on board peripherals...

Страница 96: ...nnected using the side connector panel Figure 3 5 Rear and left Side Connector Panel 1 Security Lock 2 Port for DC In AC Adapter 3 Composite Out Port 4 S Video Out Port 5 HDMI Port 6 Ethernet Port 7 I...

Страница 97: ...ations shall comply with the valid version of the system standard IEC60601 1 Everybody who connects additional equipment to the signal input part or signal output part of LOGIQ V2 LOGIQ V1 configures...

Страница 98: ...cribes how to configure the LOGIQ V2 LOGIQ V1 LOGIQ V2 LOGIQ V1 configuration For complete instructions refer to the latest revision of the LOGIQ V2 LOGIQ V1 Basic User Manual Chapter 16 Information i...

Страница 99: ...ter B W USB Printer USB port Sony UP D898MD Printer Color USB Printer USB port Sony UP D25MD Printer Officejet 100 Printer USB port HP Officejet 100 Printer Officejet Pro 8100 Printer USB port HP Offi...

Страница 100: ...Digital Printer Table 3 5 Materials furnished with B W Printer Item Description Quantity Note 1 Sony UP D897 Printer 1 2 Sony UP D898MD Printer 1 3 Paper Roll 1 4 USB cable 1 Table 3 6 Materials furni...

Страница 101: ...3 8 Materials furnished with USB Stick Item Description Quantity Note 1 Keeber USB Flash Drive 8G 1 2 Paper Roll 1 3 USB cable 1 Table 3 9 Materials furnished with the USB HDD Item Description Quantit...

Страница 102: ...d Manpower One person 5 min Preparations 1 Unpack the Sony UP D25MD Printer Installation Procedure 1 Place the device in a suitable place 2 Connect the USB Cable and power cable on the Printer 3 Conne...

Страница 103: ...7 Connectivity Service 7 Highlight Standard Print in the Service list Select the printer from the Printer pull down Properties menu Type the printer name in the Name Filed This name is used on the But...

Страница 104: ...Sony UP D25MD Printer Installation continued 8 Select Button Select the appropriate print key Print Store Print3 from the Physical Print Buttons selection Select the printer from MyComputer column an...

Страница 105: ...pack the HP Officejet 100 Printer Installation Procedure 1 Place the device in a suitable place 2 Connect the USB Cable on the Printer 3 Connect the power cord with the AC output in the wall outlet th...

Страница 106: ...HP Officejet Pro 8100 Printer Installation Procedure 1 Place the device in a suitable place 2 Connect the USB Cable on the Printer 3 Connect the power cord with the AC output in the wall outlet then t...

Страница 107: ...1 Unpack the UP D897 Printer Installation Procedure 1 Place the device in a suitable place 2 Connect the USB Cable on the Printer 3 Connect the power cord with the AC output in the wall outlet then tu...

Страница 108: ...ck the UP D898MD Printer Installation Procedure 1 Place the device in a suitable place 2 Connect the USB Cable on the Printer 3 Connect the power cord with the AC output in the wall outlet then turn o...

Страница 109: ...v 10 Footswitch Installation Tools No special tools needed Manpower One person 2 min Preparations 1 Unpack the Footswitch 2 Ensure no physical damage Installation Procedure 1 Connect the Footswitch to...

Страница 110: ...ng Footswitch Footswitch supports these configurations No Function Freeze Next Heartcycle Prev Heartcycle Print1 Print2 Update Next Step Scan Assistant Previous Step Scan Assistant Scan Assistant paus...

Страница 111: ...glish Rev 10 USB HDD Installation Tools No special tools needed Manpower One person 1 min Preparations 1 Unpack the USB HDD 2 Ensure no physical damage Installation Procedure 1 Connect the USB HDD to...

Страница 112: ...Installation Tools No special tools needed Manpower One person 1 min Preparations 1 Unpack the Wireless LAN Adapter 2 Ensure no physical damage Installation Procedure 1 Connect the Wireless LAN Adapte...

Страница 113: ...100 English Rev 10 ECG Installation Tools No special tools needed Manpower One person 1 min Preparations 1 Unpack the ECG 2 Ensure no physical damage Installation Procedure 1 Connect the ECG to the US...

Страница 114: ...y setup TCP IP Screen 1 Press Utility on the control panel and login as admin refer to Logging on to LOGIQ V2 LOGIQ V1 as ADM on page 4 11 2 Select Connectivity on the screen 3 Select TCP IP tab the s...

Страница 115: ...610739 100 English Rev 10 Changing the AE title and or Port Number To change AE Title and or Port Number edit the respective fields Figure 3 20 AE Title Port No Network setup For network connection se...

Страница 116: ...r port number The DICOM server AE title the server application s name Table 3 11 Utility Connectivity TCP IP screen 1 Enter the LOGIQ V2 LOGIQ V1 computer name This may be the same as the station name...

Страница 117: ...the LOGIQ V2 LOGIQ V1 AE stands for Application Entity DICOM services use this to identify the LOGIQ V2 LOGIQ V1 AE title is case sensitive This title may contain the Computer Name from the TCP IP pa...

Страница 118: ...ct Add The system creates a device called NewDevice 2 Change the name to one of your choosing 3 Enter the IP address of the device 4 Save your changes and then press Ping A Smiley Face indicates succe...

Страница 119: ...have already setup the device in the Device tab in How to Setup and Use a DICOM Image Storage Service on page 3 39 2 From the Select Service Type to add drop down menu select DICOM Image Storage and...

Страница 120: ...ase contact with your local GE Ultrasound representative to get the latest information 1 Power on the system NOTE Keep the power cord connection during the installation 2 After the power up sequence i...

Страница 121: ...is no need to restart the system after each installation if several option keys are installed at one time Select Cancel for the first several times and select OK after the last installation to activa...

Страница 122: ...the operation and service of the complete Ultrasound system Contents in this Section User s Manual s on page 3 42 Product Locator Installation Card on page 3 42 User s Manual s User Check that the co...

Страница 123: ...1 5610739 100 English Rev 10 Chapter 4 General Procedures and Functional Checks This chapter provides procedures for quickly checking major functions of the LOGIQ V2 LOGIQ V1 and diagnostics instruct...

Страница 124: ...functions of the scanner and diagnostics instructions using the built in service software Contents in this chapter Overview on page 4 2 General procedures on page 4 3 Functional checks on page 4 24 Po...

Страница 125: ...and cooling purposes WARNING Energy Control and Power Lockout for LOGIQ V2 LOGIQ V1 When servicing parts of the Ultrasound system where there is exposure to voltage greater than 30 volts 1 Follow LOCK...

Страница 126: ...cedures in one place Contents in this section Power ON Boot Up on page 4 5 Power off on page 4 9 Logging on to LOGIQ V2 LOGIQ V1 as ADM on page 4 11 Service Key Dongle HASP on page 4 12 Exit to Window...

Страница 127: ...DEFEATS THE SAFETY GROUND DANGER ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT THE POWER PLUG IS THE PROPER HOSPITAL GRADE TYPE WHERE REQUIRED CAUTION Ultrasound system requires all covers O...

Страница 128: ...ug the AC adapter output connector into the system DC input port located on the system s rear panel with the arrow side upward Plug the AC adapter power cord into a grounded protective earth outlet Af...

Страница 129: ...listed in the next steps d Back End Processor is turned ON and starts to load the software e The Start Screen is displayed on the monitor f A start up bar indicating the time used for software loading...

Страница 130: ...s been changed by 24 hours earlier than the current system time of last boot up This warning message is to remind the user to check the system date in case the system date and time is incorrect Figure...

Страница 131: ...g off the current user The system remains ON and ready for a new user to log on If the Logoff button is dimmed it indicates that no user is logged on to the unit at the moment Shutdown Use this button...

Страница 132: ...ect this button when you want to exit to Windows Desktop Cancel Use this button to exit from the System Exit menu and return to the previous operation System shutdown Disconnect the Mains Power Cable...

Страница 133: ...rd login Cancel Cancel login As default two users are defined USR and ADM USR If you log on as USR you will have access to do set up tasks that a user may need to do during daily use ADM If you log on...

Страница 134: ...100 English Rev 10 Service Key Dongle HASP A Service Dongle and a proprietary GE Service password are necessary for use by GE Service when performing proprietary level diagnostics like accessing the d...

Страница 135: ...LOGIQ V1 application software Follow these steps if you want to exit to the Windows Desktop when the application software is running The Service Dongle must be plugged in 1 On the alphanumeric keyboa...

Страница 136: ...to Windows Desktop from the LOGIQ V2 LOGIQ V1 application software continued 2 Select EchoLoader exe in the Process list then select End Process Figure 4 7 Maintenance Access 3 Select Yes to continue...

Страница 137: ...ure 4 9 Maintenance Access NOTE READ THIS NOTE BEFORE YOU CONTINUE After you have selected OK the Start Application dialog will be displayed on screen Be ready to select MAINTENANCE before the Time Ba...

Страница 138: ...tinued 5 Select MAINTENANCE to enter maintenance mode This will bring up the Maintenance dialog Figure 4 10 System halted 6 In the Maintenance window select Exit to Windows This will display the Windo...

Страница 139: ...Service Manual 4 17 5610739 100 English Rev 10 Removable media Refer to the latest revision of the User Manual to perform the following tasks Using Removable Media Labeling Removable Media Formatting...

Страница 140: ...d on an DVD R disk or USB memory device for reloading on the system NOTE Presets should NOT be saved on the same USB memory device or DVD R disk as images The Archive Menu lists the images but does NO...

Страница 141: ...he destination 5 Select Backup The backup status for each item is displayed on the Result column Loading Presets from a USB memory device or DVD R 1 Connect the USB memory device or DVD R with the arc...

Страница 142: ...anagement Schedule Configuring Data Management Settings Configuringestination Device Setting Running the Disk Management Function Starting Disk Management Manually Backup For more information refer to...

Страница 143: ...PDF files from the Manual CD 1 Insert the CD R disk CD ROM into the CD drive on a PC or Laptop with Adobe Acrobat Reader 2 Follow the instructions on the screen to display the manual of choice 3 Befor...

Страница 144: ...eps 1 Power down the system 2 Disconnect the mains power cable from the wall outlet Follow these links if you need more information See Power off on page 4 9 for more information WARNING DO NOT touch...

Страница 145: ...e dust gasket counterclockwise until it can be removed from the keyboard Figure 4 13 Remove the retainer 3 Separate the trackball and the gasket Wipe off any oil or dust from the trackball gasket and...

Страница 146: ...section the functional checks for LOGIQ V2 LOGIQ V1 are described Functional checks are used to verify that the product works as intended Functional checks may also be used during troubleshooting Cont...

Страница 147: ...ard 6 User Defined keys 7 Report key 8 Utility key 9 Patient key 10 Preset Key 11 Worksheet key 12 End Exam key 13 Archive key 14 Gain AO keys 15 Scan Coach keys 16 Mode keys 17 Cursor key 18 Clear ke...

Страница 148: ...d Functional Checks 4 26 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Performance Tests Test Phantoms The use of test phantoms is only recommended if required by your facility s c...

Страница 149: ...ust the range to be imaged Use Focus to center the focal point s around the region of interest Use Frequency move to higher frequencies or Frame rate move to lower frame rate to increase resolution in...

Страница 150: ...rify that all the listed controls are working as intended Adjust Horizontal sweep to optimize the display resolution Adjust Gain and TGC controls to adjust the range to be imaged Use the Frequency mov...

Страница 151: ...from M mode or a combination of these Preparations Connect one of the probes to the scanner s left most probe connector Turn ON the scanner The B Mode window is displayed default mode Select Color Mo...

Страница 152: ...Lower settings improve depth penetration and sensitivity This does not affect the frequency used for 2D and M Mode NOTE Frequency setting may affect FPS SV and Low Velocity Reject Adjust Power to obta...

Страница 153: ...100 English Rev 10 Select Color M Mode 1 Select M Mode 2 Use the trackball to position the ROI frame over the area to be examined 3 Press Select The instruction Size should be highlighted in the trac...

Страница 154: ...ower settings improve depth penetration and sensitivity This does not affect the frequency used for 2D and M Mode NOTE NOTE Frequency setting may affect FPS SV and Low Velocity Reject Adjust Power to...

Страница 155: ...robe or with an ordinary probe By using an ordinary probe you can first bring up a 2D picture for navigation purpose and then add PW Doppler Preparations Connect one of the probes to the scanner Turn...

Страница 156: ...Frequency to optimize flow display Higher setting will improve resolution and the lower setting will increase the depth penetration Adjust Frame rate to a higher setting to improve motion detection o...

Страница 157: ...es with high temporal resolution Preparations Connect one of the probes to the scanner s left most probe connector Turn ON the scanner The 2D Mode window is displayed default mode If needed adjust the...

Страница 158: ...wever from this window the beam cannot be aligned to the muscle for all the parts of the ventricle NOTE PW will be optimized for Tissue Velocities when activated from inside TVI Basic Measurements NOT...

Страница 159: ...be head could result in irreparable damage Table 4 1 Probe and Connectors Checks Step Task Expected Results 1 Select the appropriate connected probe from the probe indicators on the Touch Panel The pr...

Страница 160: ...connecting the probes For available probes see Probe and Accessories on page 9 9 Turn ON the scanner The 2D Mode is displayed default mode Adjust the Cineloop controls Press Freeze The left and right...

Страница 161: ...Panel Test The Operator Panel is tested when the LOGIQ V2 LOGIQ V1 is powered up as part of the start up scripts run at every start up For more info please refer to Diagnostics Troubleshooting on pag...

Страница 162: ...ower Supplies if there don t seems to be a problem that may be related to the Power Supply Power Supply Adjustment There are no adjustments on the power supply The DC Power is self regulated If a volt...

Страница 163: ...ing the scanner to customer for use Software Configuration Checks Table 4 2 Software Configuration Checks Step Task to do Notes 1 Verify Date and Time is correct 2 Verify that Location Hospital Name i...

Страница 164: ...General Procedures and Functional Checks 4 42 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Site Log Table 4 3 Site Log DATE SRVICE PERSON PROBLEM COMMENTS...

Страница 165: ...9 100 English Rev 10 Chapter 5 Components and Functions Theory This chapter explains LOGIQ V2 LOGIQ V1 s system concepts component arrangement and subsystem functions It also describes the power distr...

Страница 166: ...ory 5 2 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Overview Contents in this chapter Overview on page 5 2 Block Diagram and Theory on page 5 3 Power Diagram on page 5 5 Common S...

Страница 167: ...ons of the above Signal flow from the Probe Connector Panel to the Front End to the Mid Processors and Back End Processor and finally to the LCD and peripherals System configuration is stored on a har...

Страница 168: ...provides multi powers to MST PWA and IO PWA MST provides the main control function 64 channels transmitting receiving and related signal processing and connects to CWD board IO board receives MISC si...

Страница 169: ...utput to the DC DC unit which is inside the system console The input of AC power pack will be the AC outlet and it s universal the range is AC 100V 240V 50 60Hz And no main power switch located on thi...

Страница 170: ...100 English Rev 10 Common Service Platform Introduction The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems The Common Service...

Страница 171: ...V1 Basic Service Manual 6 1 5610739 100 English Rev 10 Chapter 6 Service Adjustments This chapter describes how to test and make adjustments to the LOGIQ V2 LOGIQ V1 You can use these to test the sys...

Страница 172: ...Service Adjustments 6 2 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Overview Contents in this chapter Overview on page 6 2 LCD Monitor adjustments on page 6 3...

Страница 173: ...ection This section describes how to test and adjust the scanner These tests are optional You may use them to check the system for errors Monitor Adjustments To adjust the brightness and volume 1 On t...

Страница 174: ...e sure that the DICOM spooler is empty Review the troubleshooting section to collect information for future analysis if you are troubleshooting loss of image or patient data Be sure all USB printers a...

Страница 175: ...39 100 English Rev 10 System Reload procedure If the system running please follow the steps below 1 Enter Utility Service 2 Select GE Service in option of User Level of GEMS Service Home page and inpu...

Страница 176: ...100 English Rev 10 System Reload procedure continued 3 Enter Utility Common Utilities Disruptive Mode select Yes Figure 6 3 Enable Disruptive Mode 4 Choose Utilities in Global Service Interface and th...

Страница 177: ...V1 Basic Service Manual 6 7 5610739 100 English Rev 10 System Reload procedure continued 5 Select Setup Software Figure 6 5 Invoke System Reload 6 Reload system setup is successful then reboot the sy...

Страница 178: ...rocedure continued 7 The software will be loading Wait for the software installation to complete Status on the screen indicates the progress Figure 6 6 System Reload 8 When the software reload process...

Страница 179: ...oftware Download Software Download setting up 1 The user receives notification of available software updates via the envelop icon that appears in the status bar on the bottom of the monitor Figure 6 7...

Страница 180: ...ne evening and then install the software the next evening After the software installation is complete you will be asked to perform a few system functional checks to determine normal system operation N...

Страница 181: ...he Exit dialog window with software download displays Figure 6 8 System Exit Dialog Window For R1 0 x Figure 6 9 System Exit Dialog Window For R1 1 x Decline You decline the software download the soft...

Страница 182: ...OGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Software download and installation continued 2 Select Download Figure 6 10 System Exit Dialog Window For R1 0 x Figure 6 11 System Exit...

Страница 183: ...ble until the software installation is done While installing DO NOT turn off system power NOTE A typical software update of about 600 Mb may take up to 60 minutes to download but times may vary depend...

Страница 184: ...omplete Select Install Figure 6 12 Software Ready To Be Installed For R1 0 x Figure 6 13 Software Ready To Be Installed For R1 1 x Decline DO NOT install the downloaded software no software upgrade wi...

Страница 185: ...allation process DO NOT interrupt this process AND follow instructions as they appear on the display NOTE A typical installation may take up to 15 minutes 4 After the installation complete select Shut...

Страница 186: ...0 Software download and installation continued 5 When the system starts up after the software installation has finished press Power On Off button any then select Verify Figure 6 16 Shutdown after Soft...

Страница 187: ...g is critical You MUST perform software verification after downloading and installing the new software Figure 6 18 New Software Verification Checklist CAUTION Perform a check for all the features list...

Страница 188: ...ev 10 Software download and installation continued 7 As you verify that each feature works correctly select Passed If all features work correctly and Passed is filled in for all features type your sig...

Страница 189: ...tinued WARNING However if any of the features DO NOT function as expected you need to select Failed for this feature Type your signature and press OK Then the System Exit Window is displayed select Ex...

Страница 190: ...Service Adjustments 6 20 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10...

Страница 191: ...5610739 100 English Rev 10 Chapter 7 Replacement Procedures This chapter describes how to remove and install or replace modules and subsystems in the LOGIQ V2 LOGIQ V1 It also includes instructions fo...

Страница 192: ...V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Overview Contents in this chapter Overview on page 7 2 Warnings and important information on page 7 3 Disassembly Re assembly on page 7 5 L...

Страница 193: ...t least 30 seconds for capacitors to discharge as there are no test points to verify isolation 6 Remove disconnect the battery if present Ultrasound System components may be energized WARNING Because...

Страница 194: ...ipped from a clinic or site for example body coils or an ultrasound probe The purpose of the regulation is to protect employees in the transportation industry as well as the people who will receive or...

Страница 195: ...WARNING ONLY QUALIFIED SERVICE PERSONNEL SHOULD REMOVE ANY COVERS OR PANELS ELECTRICAL HAZARDS EXISTS AT SEVERAL POINTS INSIDE BECOME THOROUGHLY FAMILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT L...

Страница 196: ...s Screwdriver 7 screw 5244775 2 Screw GB T820 2000 M3x8 Common Phillips Screwdriver 8 screw 5342274 4 Inch SScrew 6 32UNC Common Phillips Screwdriver 9 screw 5439265 246 Screw M4x10 Common Phillips Sc...

Страница 197: ...Rotate the retainer 2 Seperate the trackball and the retainer Wipe off any oil or dust from the trackball retainer and the trackball housing using a cleaner or cotton swab 3 Assemble the trackball and...

Страница 198: ...5610739 100 English Rev 10 2 Probe Holder 5727508 Purpose This is a description on how to remove and replace the 2 Probe Holder Tools Common Phillips screwdrivers Preparation Lock the castors in orde...

Страница 199: ...er 5727508 continued Removal Procedure Mounting Procedure 1 Install the new parts in the reverse order of removal Table 7 2 Removal Procedure for 2 Probe Holder No Steps Corresponding Graphic 1 Unscre...

Страница 200: ...y Kit of V Cart Tools None Preparation Lock the castors in order to avoid cart moving Needed Manpower 1 person 1 minutes travel Removal Procedure Mounting Procedure 1 Install the new parts in the reve...

Страница 201: ...10739 100 English Rev 10 Printer Shelf 5727506 Purpose This is a description on how to remove and replace the Printer Shelf of V Cart Tools Common Phillips screwdrivers Preparation Lock the castors in...

Страница 202: ...d Removal Procedure Mounting Procedure 1 Install the new parts in the reverse order of removal Table 7 4 Removal Procedure for Printer Shelf No Steps Corresponding Graphic 1 Remove the screw pad 2 Uns...

Страница 203: ...mages to a remote server or to removable media For instructions please see Disk management in the User Manual User Guide CAUTION An error or a power loss may occur Always backup the Patient Archive an...

Страница 204: ...rom the following screens CAUTION An error or a power loss may occur Always backup the Patient Archive and the Presets System Configurations before loading the software In order to complete a successf...

Страница 205: ...stick Load the system software with SD Card NOTE Before starting this procedure remove all probes and peripherals and remove them from the Docking Cart NOTE While it is believed to be unnecessary it w...

Страница 206: ...re to the system 2 Properly turn off the scanner by momentarily pressing the Power On Off Switch In System Exit window select Shutdown to shutdown the system NOTE If the system will not shutdown norma...

Страница 207: ...recovered after wipe process To select 3 Service tool The complete disk will be loaded This option is recommended for application software upgrade NOTE This option is used by service engineer only It...

Страница 208: ...10739 100 English Rev 10 Loading the System Software with USB memory stick For R1 0 x continued 5 System USB memory stick will be loading Wait for the software installation to complete Figure 7 3 Load...

Страница 209: ...0 x continued 6 Select Finish when the upgrade process completes and the system will shutdown automatically Figure 7 4 System upgrade complete 7 Remove the USB memory stick and then press Power On Of...

Страница 210: ...that of service dongle Figure 7 5 Input password For software R1 0 6 and above the following dialog will display There are two options for you to continue If you will input a new serial number in ste...

Страница 211: ...9 Then input the correct Serial number select OK NOTE The serial number is case sensitivity Figure 7 7 Input Serial Number NOTE The serial number can be found at the bottom of the system 10 The seria...

Страница 212: ...GIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Loading the System Software with USB memory stick For R1 0 x continued 11 Power on the system input the option keys if the dialog displa...

Страница 213: ...ta on the disk has been backed up NOTE Patient Archive Report Archive User Defined Configuration and Service of R1 0 x are not compatible with R1 1 x Do not restore them between R1 0 x and R1 1 x BIOS...

Страница 214: ...0739 100 English Rev 10 Loading the System Software with USB memory stick For R1 1 x continued 2 Input the BIOS password gerural and login the BIOS Figure 7 11 BIOS Password 3 In the BIOS Main page se...

Страница 215: ...h USB memory stick For R1 1 x continued 4 Check the CPU GPN number recorded on the step 1 on page 7 23 and select AHCI for SATA Mode There are two setting processes to continue If the GPN number is 55...

Страница 216: ...tem Software with USB memory stick For R1 1 x continued If the GPN number is 5505176 select AHCI via the Chipset PCH IO Configuration SATA Configuration SATA Mode Selection Press F10 to save and exit...

Страница 217: ...t software upgrade procedure 1 Insert the Upgrade USB memory stick labeled System Application Software in the USB port of the system 2 Properly turn off the system by momentarily pressing the Power On...

Страница 218: ...ition only this option is intended for recovery of a system that will not boot up All patient data is preserved This option is not applicable for software upgrade from R1 0 x to R1 1 x NOTE While the...

Страница 219: ...If the Load the complete disk is selected For R1 0 x read the information in the Confirmation dialogue box carefully and check the boxes before Preset Patient data and BIOS to confirm that the require...

Страница 220: ...cedure For R1 1 x continued For R1 1 x input the Installation Password GEWuXiChina Then select Execute to continue Figure 7 17 Input Password between R1 1 x If Load the bootable C partition only is se...

Страница 221: ...ate the SSD Firmware and shut down afterwards Restart the system by pressing Power On Off key on the control panel and reselect Load the complete disk If no the system will automatically continue runn...

Страница 222: ...10739 100 English Rev 10 Software Upgrade Procedure For R1 1 x continued 8 The system will display upgrading result and shutdown automatically within 30 seconds Or click the Shutdown in the screen to...

Страница 223: ...mory stick the software system upgrading process repeats when the system boots up NOTE Ensure the USB memory stick is properly and securely connected Once the USB memory stick is accidentally disconne...

Страница 224: ...place the original one NOTE The Windows License Label to replace the original one is only required for upgrade from R1 0 x to R1 1 x 1 Take out the windows license label pell off the yellow part of th...

Страница 225: ...ash drive labeled System Application Software to the system 2 Properly turn off the system by momentarily pressing the Power On Off Switch In System Exit window select Shutdown to shutdown the system...

Страница 226: ...then select Next to continue For R1 1 x software select Wipe tool NOTE All patient data cannot be recovered after this operation Contact GE service for support NOTE All patient data if any will be de...

Страница 227: ...pe Tool continued 5 The system indicates all data will be erased select Yes to continue Figure 7 26 Confirmation Dialog For R1 0 x Figure 7 27 Confirmation Dialog For R1 1 x NOTE All the patient data...

Страница 228: ...10739 100 English Rev 10 Wipe Tool continued 6 The process may take twenty minutes or more NOTE Do not interrupt the wipe tool process at any time until all the partitions wiping is completed Figure 7...

Страница 229: ...wn automatically when the wiping progress completes Figure 7 30 Process Complete NOTE As the SSD is empty after wipe process the system cannot boot up The software should be loaded first after the wip...

Страница 230: ...ersion check out Functional Check out 1 Power on LOGIQ V2 LOGIQ V1 system and wait until system booting to main screen 2 Press Utility on the control panel 3 Select the About button on the right Figur...

Страница 231: ...eload the Correct Preset Region NOTE After the system software loading completion please reload the correct preset region 1 Reboot the system 2 Select Utility General 3 In the Location portion select...

Страница 232: ...e activated and working 1 Reboot the system 2 Select Utility Admin System Admin 3 Ensure that all the installed option keys are displayed and the status of Options are valid The status Valid means the...

Страница 233: ...on is displayed on the right top location of the screen About the probe specification for intended use on LOGIQ V2 LOGIQ V1 Plug in at least one of each type of the probes and check if each of the pro...

Страница 234: ...ny DICOM devices used by the customers and check to ensure these DICOM devices work properly The instruction about installing DICOM devices is not incorporated in this manual To access the instruction...

Страница 235: ...tics Troubleshooting This chapter describes LOGIQ V2 LOGIQ V1how to setup and run the tools and software that help maintain image quality and system operation Very basic host system and board levels a...

Страница 236: ...ng 8 2 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Overview Contents in this chapter Overview on page 8 2 Gathering Trouble Data on page 8 3 Screen Capture on page 8 5 Network Co...

Страница 237: ...cturer for analysis There are different options to acquire this data that would give different results Collect Vital System Information The following information is necessary in order to properly anal...

Страница 238: ...em presets and several log files in a date and time stamped zip file NOTE This function may also be used to make a Print Screen This Alt D function is available at all times When Alt D is pressed a me...

Страница 239: ...the clipboard using a Print Key Check the function of the Print Key in the event that the customer may have made some custom settings 1 Press Utility on the Control Panel 2 Select Connectivity from th...

Страница 240: ...e Setting the Store Key to Screen Capture If the Print Key is not set to screen capture 1 While on the Connectivity screen with the Button tab displayed go to the Destinations list 2 From the list sel...

Страница 241: ...Screen The following is a generic process to capture any screen from the scanner 1 Navigate to and display the image screen to be captured 2 Press Store This will place a snapshot of the screen on th...

Страница 242: ...he snapshot to be stored 4 Select Save As Icon on the right side of the image screen Figure 8 5 Save As Icon 5 A Save dialog box will be opened Choose the archive location to save image on the USB Dri...

Страница 243: ...the Store Key to Screen Capture on page 8 6 Refer to Figure 8 2 on page 8 5 1 Press Utility on the control panel 2 Select Connectivity in the Utility menu 3 In the Button field select Store 4 In the D...

Страница 244: ...em Also select to retain Disruptive Mode or Not After submitting to restart or shutdown a confirmation screen gives one last chance to confirm or cancel the request Remote access Remote access is a fe...

Страница 245: ...Diagnostics LOGIQ V2 LOGIQ V1 Basic Service Manual 8 11 5610739 100 English Rev 10 Utilities continued 2 Select the CSD to Connect Then the OLE is remotely connected to the system Figure 8 8 Remotely...

Страница 246: ...g 8 12 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Utilities continued 3 Enter Utility Common Utilities Disruptive Mode select Yes Figure 8 9 Enable Disruptive Mode Figure 8 10 D...

Страница 247: ...ice Manual 8 13 5610739 100 English Rev 10 Utilities continued 4 If we want to see the Scan screen or have control of the unit we have to go to Virtual Console Observation and Start it Then connect us...

Страница 248: ...oting 8 14 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Utilities continued 5 Select the VNC to Connect Then the OLE is remotely connected to the system Figure 8 12 Virtual Consol...

Страница 249: ...dulator FPGA on the GFW GFW Front End Interface Test tests that the GFW can access Front End boards GFW Analog Test GFS Memory Access Test tests that GFW can access to the internal external external c...

Страница 250: ...P Test tests HV STOP mechanism and checks each board is able to assert HV_STOP using its HV_STOP source Memory Callisto Memory Test NOTE The Callisto Memory test may fail if it is performed with other...

Страница 251: ...Common Diagnostics LOGIQ V2 LOGIQ V1 Basic Service Manual 8 17 5610739 100 English Rev 10 Doppler Tests Doppler Audio Test runs the tests for Doppler...

Страница 252: ...uccessful you will see a brief video clip with audio For more information about the test click More Information Click Pass if the test successfully reproduces the video clip Click Fail if the test is...

Страница 253: ...itor color depth capacity If the colors in the color spectrum do no blend smoothly together a problem may exist with the monitor color depth Use the graduated horizontal and vertical alignment bars to...

Страница 254: ...ection the wheel is being rotated Clicking the wheel will flash the picture of the mouse in the Mouse Status Test area The Drag and Drop Test verifies a mouse can successfully perform drag and drop op...

Страница 255: ...probe and unwrap the probe cord 3 Put the probe in the probe holder 4 Align the connector with the probe port and carefully push into place 5 Lock the probe latch upward 6 Carefully position the probe...

Страница 256: ...English Rev 10 Assessment Utility continued 8 In the Diagnostic Service Diagnostics Assessment Utility Probe Assessment to check the probe If there is no broken element and the probe is acceptable fo...

Страница 257: ...lish Rev 10 Assessment Utility continued If there are broken elements and the probe is not acceptable for diagnostic The assessment is fail and the failed elements are shown in the graphic Figure 8 14...

Страница 258: ...Basic Service Manual 5610739 100 English Rev 10 Probe Pass Fail Criteria Pass No dead element Dead elements on the edge 0 1 or end 1 end Fail Max 2 weak elements on the scan line 3dB elements 2dB Any...

Страница 259: ...LOGIQ V2 LOGIQ V1 Basic Service Manual 8 25 5610739 100 English Rev 10 OP Panel Interface Figure 8 15 OP Panel Test Restart the system after diagnostics Always shutdown the system and reboot after a d...

Страница 260: ...vice Manual 5610739 100 English Rev 10 Network Configuration Wire LAN Network 1 Connect system with network 2 Enter Utility Connectivity TCP IP in IP settings window check Enable DHCP and select the p...

Страница 261: ...on input static address in IP address Subnet Mask and Default Gateway box In Network Speed box choose the proper speed available Figure 8 17 Input Static IP 3 Select Save and a popup window displays S...

Страница 262: ...h Rev 10 Wireless LAN Network NOTE To configure the Wireless LAN network the operator must login as administrator 1 Connect the wireless adapter in the USB port 2 Enter Utility Connectivity TCP IP In...

Страница 263: ...LOGIQ V2 LOGIQ V1 Basic Service Manual 9 1 5610739 100 English Rev 10 Chapter 9 Renewal Parts This chapter lists the renewal parts available for the LOGIQ V2 LOGIQ V1...

Страница 264: ...Renewal Parts 9 2 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 Overview Contents in this chapter List of Abbreviations Renewal Parts Lists...

Страница 265: ...eviations Table 9 1 List of Abbreviations ABBREVIATION DESCRIPTION 3D THREE DIMENSIONAL Assy ASSEMBLY FRU 1 Replacement part available in part hub FRU 2 Replacement part available from the manufacture...

Страница 266: ...176304 2 AC Power Cord China 1 1 003 5176773 2 AC Power Cord India 1 1 004 5177195 2 AC Power Cord Argentina 1 1 005 5176907 2 AC Power Cord UK 1 1 006 5177153 2 AC Power Cord Denmark 1 1 007 5177154...

Страница 267: ...Renewal Parts Lists LOGIQ V2 LOGIQ V1 Basic Service Manual 9 5 5610739 100 English Rev 10 Operator Console Assy Figure 9 1 Operator Console Assy V Cart Figure 9 2 V Cart...

Страница 268: ...h Rev 10 V Cart Table 9 3 V Cart Item Part Number Part Name Corresponding graphic Qty FRU 1400 5732311 S V Cart SVC Parts 1 1 5720204 V Cart Front Handle 5720225 V Cart Handle Clip 5720207 V Cart Cabl...

Страница 269: ...it 1 1 1402 5727506 V Cart Printer Shelf 1 1 1403 5727508 V Cart Probe Holder For 2PP and Cart 1 1 1404 5725649 S V Cart Caster Kits 2 1 5725649 2S V Cart Castor Kits 2 1 1405 5730407 S V Cart Bigger...

Страница 270: ...l Parts 9 8 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10 1407 5730407 2S Bigger Castor kit including Wrench 1 1 Table 9 3 V Cart Item Part Number Part Name Corresponding graphic Q...

Страница 271: ...robe and Accessories Table 9 4 Probes for LOGIQ V2 LOGIQ V1 Item Part Name Part Number Center Image Frequency MHz Qty FRU 700 4C RS 5451471 3 10 1 1 701 L6 12 RS 5454332 7 75 1 1 702 3Sc RS 5433833 2...

Страница 272: ...5151262 UP D897MD B W Printer Chinese kit 1 816 5151259 UP D897MD B W Printer USA kit 1 817 5151261 UP D897MD B W Printer European kit 1 818 5151263 UP D897MD B W Printer Japanese kit 1 819 5426594 HP...

Страница 273: ...ystem and Application Software USB 1 834 5765267 2S LOGIQ V2 LOGIQ V1 R1 1 1 System and Application Software USB 1 ECG 835 H48952AB ECG cable kits EU 1 836 H41852LL ECG IEC Lead 1 837 H41852LK ECG AHA...

Страница 274: ...ia 1 1 903 5177195 2 AC Power Cord Argentina 1 1 904 5176907 2 AC Power Cord UK 1 1 905 5177153 2 AC Power Cord Denmark 1 1 906 5177154 2 AC Power Cord Switzerland 1 1 907 5177187 3 AC Power Cord Aust...

Страница 275: ...version 1 N 1011 5610736 129 LOGIQ V2 LOGIQ V1 User Guide Estonian version 1 N 1012 5610736 131 LOGIQ V2 LOGIQ V1 User Guide Slovenian version 1 N 1013 5610736 140 LOGIQ V2 LOGIQ V1 User Guide Japanes...

Страница 276: ...030 5610736 174 LOGIQ V2 LOGIQ V1 User Guide Lithuanian version 1 N 1031 5610736 175 LOGIQ V2 LOGIQ V1 User Guide Latvian version 1 N 1032 5610736 176 LOGIQ V2 LOGIQ V1 User Guide Serbian version 1 N...

Страница 277: ...Maintenance This chapter describes Care and Maintenance on the Ultrasound system and peripherals These procedures are intended to maintain the quality of the Ultrasound system s performance Read this...

Страница 278: ...on page 10 2 Warnings on page 10 3 Why do maintenance on page 10 4 Maintenance task schedule on page 10 6 Tools required on page 10 8 System maintenance on page 10 12 Electrical safety tests on page...

Страница 279: ...AUTIOUS WHENEVER POWER IS STILL ON AND COVERS ARE REMOVED CAUTION Practice good ESD prevention Wear an anti static strap when handling electronic parts and even when disconnecting connecting cables CA...

Страница 280: ...s Quality Assurance Programs may require additional tasks and or inspections at a different frequency than listed in this manual Keeping records It is good business practice that ultrasound facilities...

Страница 281: ...hysicists the supervising radiologist physician or appropriate designee Routine quality control testing must occur regularly The same tests are performed during each period so that changes can be moni...

Страница 282: ...r LOGIQ V2 LOGIQ V1 and can best provide competent efficient service Contact GE for coverage information and or price for service The service procedures and recommended intervals shown in the Care and...

Страница 283: ...e Annually Peripheral Leakage Current Checks See Notes Twice Annually Surface Probe Leakage Current Checks See Notes Twice Annually Endocavity Probe Leakage Current Checks See Notes Quarterly Annually...

Страница 284: ...n End 1 2 9 16 9 XL9974MM Xcelite hex Blade 2 5mm 9 44579 Wrench Open End 1 2 9 16 9 XL9975MM Xcelite hex Blade 3mm 9 45385 Pliers Arc Joint 7 inch 9 XL9976MM Xcelite hex Blade 4mm 9 45378 Pliers Slip...

Страница 285: ...lade 5 32 inch 9 43499 1 2 inch Socket 1 4 inch drive 9 XL9926 Xcelite hex Blade 3 16 inch 9 4355 Flex Spinner 9 XL99764 Xcelite hex Blade 7 64 9 43523 Breaker 9 XL99964 Xcelite hex Blade 9 64 9 43531...

Страница 286: ...Punch Pin 1 4 inch 9 34374 3 8 inch Metric Socket Set 12 PT 9 42973 Cold Chisel 1 2 inch 9 44311 16mm Socket 12 pt 9 GH77 Center Punch Automatic 9 33485 Metal Socket Tray 9 GH890 File Handle Adj 9 334...

Страница 287: ...trap 3M 2214 Small adjustable wrist strap 3M 3051 conductive ground cord Anti Static Vacuum Cleaner 120V 230V Safety Analyzer The safety Analyzer tool should be calibrated and compliant with AAMI ESI...

Страница 288: ...rasound system power on and verify that all fans and peripherals turn on Watch the displays during power up to verify that no warning or error messages are displayed Where applicable confirm that the...

Страница 289: ...be Elements Perform an Element Test on each probe to verify that all the probe elements and system channels are functional 6 Applicable Software Options Verify the basic operation of all optional mode...

Страница 290: ...sary 3 Color Printer Verify hardcopy output of the Color video page printer Clean heads and covers if necessary 4 DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 5 ECG...

Страница 291: ...r for dents scratches or cracks 3 Input Power Refer to Mains cable inspection on page 10 14 4 Cables Connectors Check all internal cable harnesses and connectors for wear and secure connector seating...

Страница 292: ...perature log to see if there are any trends that could cause problems in the future 3 Check the Configuration Log update if needed 10 Power and System Status Indicators Check for proper operation of a...

Страница 293: ...Failure to properly handle or maintain a probe may also void its warranty Any evidence of wear indicates the probe cannot be used Do a visual check of the probe pins and Ultrasound system sockets befo...

Страница 294: ...de of nitrile derived from vegetable starch to prevent allergic latex reactions CAUTION Failure to follow the prescribed cleaning or disinfection procedures will void the probe s warranty DO NOT soak...

Страница 295: ...low chart below to carry out battery performance maintenance Figure 10 1 Flow chart of Battery Performance Maintenance NOTE Disconnect all probes when discharge battery NOTE Discharge the battery to l...

Страница 296: ...THS ACCORDING TO THE HISTORICAL DATA ONLY TRAINED PERSONS ARE ALLOWED TO PERFORM THE SAFETY INSPECTIONS MENTIONED ABOVE DANGER TO MINIMIZE RISK OF ELECTRICAL SHOCK ONLY TRAINED PERSONS ARE ALLOWED TO...

Страница 297: ...e cables and strain relief at each end Inspect the transducer enclosure and lens for cracks holes and similar defects Test the system peripherals and probes for leakage current Excessive leakage curre...

Страница 298: ...for capacitors to discharge as there are no test points to verify isolation 6 Remove disconnect the battery if present Ultrasound System components may be energized CAUTION Compare all safety test res...

Страница 299: ...able 10 14 Type CF Applied Part Leakage Current Limits ECG Connections Country Normal Condition Open Ground Reverse Polarity Open Neutral Mains Applied All 0 1 mA 0 5 mA 0 5 mA 0 5 mA 5 0 mA Table 10...

Страница 300: ...be reported to the hospital immediately and the receptacle should not be used Figure 10 2 Typical alternate outlet tester NOTE No outlet tester can detect the condition where the Neutral grounded supp...

Страница 301: ...xposed metal parts of the case The ground wire resistance should be less than 0 2 ohms Reference the procedure in the IEC60601 1 Figure 10 3 Ground continuity test DANGER ELECTRIC SHOCK HAZARD THE PAT...

Страница 302: ...lar test instrument its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter This places the meter in the grounding con...

Страница 303: ...n page 10 22 Record all data on the PM Inspection Certificate Table 10 16 Typical Data Sheet for enclosure Source Leakage Current Unit Power Tester Polarity Switch Tester Neutral or Ground Switch Test...

Страница 304: ...e and test lead placement It is abnormal if no leakage current is measured If no leakage current is detected check the configuration of the test equipment Tools For needed tools see Tools required on...

Страница 305: ...er Procedure Using Probe Adapter Follow the Safety Analyzer tool instruction to test each transducer for leakage current The electrical Safety Analyzer tool should be calibrated and compliant with AAM...

Страница 306: ...ss than the values shown in Table 10 13 on page 10 22 Record all data on the PM Inspection Certificate Table 10 17 Typical Data Sheet For Transducer Source Leakage Current Transducer Tested Unit Power...

Страница 307: ...Where applicable tighten all grounds Ensure star washers are under all ground studs Inspect wiring for bad crimps poor connections or damage Test the wall outlet verify it is grounded and is free of...

Страница 308: ...obe type in the appropriate space on the check list If excessive leakage current is slot dependent inspect the system connector for bent pins poor connections and ground continuity If the problem rema...

Страница 309: ...N r e m o t s u C Date Performed Warranty C ontract HBS System Type e t a D e r u t c a f u n a M r e b m u N l a i r e S r e b m u N l e d o M r e b m u N l a i r e S r e b m u N l e d o M t a m r o...

Страница 310: ...o D O I l a n r e t x E n o i t c n u F e d o M F C s r o t c e n n o C d n a s e l b a C n o i t c n u F e d o M M Applicable Software Opti l a r e h p i r e P d e v o r p p A E G s n o s DVD RW Prin...

Страница 311: ...lue measured OK Comments Outlet correct ground and wiring config System ground continuity Chassis source leakage current probe Chassis source leakage current wheel Chassis source leakage current monit...

Страница 312: ...2 ground continuity Peripheral 3 leakage current Peripheral 3 ground continuity Table 10 18 Electrical safety tests log Continued Electrical test performed Max value allowed Value measured OK Comment...

Страница 313: ...rs 1 28 customer order verification 3 13 customer provided prerequisite 7 13 D damage in transportation 3 7 damage in transport i 14 dangerous procedure warnings 1 22 data network setup requirements 2...

Страница 314: ...ents 2 19 installation see system setup 3 1 installation warnings see setup warnings 3 3 introduction chapter 1 1 1 L legal notes i 17 lighting 2 5 list of abbrevations 9 3 loading software 7 13 locko...

Страница 315: ...site preparations 2 1 time and manpower requirements 2 11 software loading 7 13 software hardware compatibility 9 4 software hardware compatibility 5 3 specifications 3 14 electrical specifications 3...

Страница 316: ...Index 4 LOGIQ V2 LOGIQ V1 Basic Service Manual 5610739 100 English Rev 10...

Страница 317: ...GE...

Отзывы: