GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
8 - 186
Section 8-11 - Replacing Options
8-11-3 Replacement of the Rear Handle Cable Hook Set-U
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-11-3-1
Removal Procedure
8-11-3-2
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-98 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
Flat Screwdriver
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1.
Release 2 snap fits to remove the Unity Handle Cable Hook L/R
from the console
Содержание LOGIQ Totus
Страница 2: ......
Страница 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Страница 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Страница 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Страница 198: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 4 32 Section 4 4 Site Log...
Страница 236: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 5 38 Section 5 14 Product Manuals...
Страница 627: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual Chapter 10 Care Maintenance 10 25...
Страница 642: ...GE HealthCare CONFIDENTIAL Direction 5936427 Revision Rev 1 LOGIQ Totus Basic Service Manual Index 4...
Страница 643: ......