Section 7-8 - Error messages
7 - 93
7-8-1
FibroScan Module Error messages and Logs
Table 7-21:
Error messages and Logs
Message: A system reboot is required to connect to the FibroScan
module.
The system has detected FibroScan Module connection is not available.
Reboot the system.
Cause: The network connection between FibroScan Module and the
system was disconnected during FibroScan mode, and connection
recovery failed.
Message: FibroScan is not available. A FibroScan module is required
The system could not detect FibroScan Module. Please check below
steps:
1. FibroScan Module is mounted on the system
2. Green LED on FibroScan Module is ON
3. Blue LED on FibroScan Module is ON
If either of above step is failed, go to trouble shooting.
Cause: The system did not detect an established network connection
between FibroScan Module and the system on boot up.
Message: Shutting down FibroScan Module, please wait...
The system will be shutting down in a moment. The system will shut down
FibroScan Module safely before shutting down itself. The system will
proceed on shut down after message disappears.
Message: FibroScan connection is not available. Reconnecting... this
could take up to 30 seconds.
The system has detected FibroScan Module connection is not available.
The system is trying to reconnect to FibroScan Module. Please simply
wait for the reconnection process to end. The message will disappear if
reconnection is successful. You could continue the FibroScan
measurement if reconnection is successful.
Cause: A network disconnection has occurred during FibroScan mode.
Message: FibroScan reconnection failed. System will exit FibroScan
mode. FibroScan data may not be saved.
The system has detected FibroScan Module connection is not available,
and reconnection process has failed. The system will force exit
FibroScan mode without storing FibroScan measurement data.
Cause: A network disconnection has occurred during FibroScan mode.
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......