GE H
EALTHCARE
D
IRECTION
5394227, 12
LOGIQ S8/LOGIQ E8 S
ERVICE
M
ANUAL
Section 7-3 - Gathering Troubleshooting Data (R1 to R4)
7 - 9
7-3-4
Collecting a Screen Capture with Logs
NOTE:
Login before gathering log files.
To gather log files, the system requires a user to be logged in. The error message below appears when
hitting Alt+D key without any user logged in.
If the system malfunctions, press the Alt+D keys simultaneously. This Alt+D function is available at all
times, and collects a screen capture of the image monitor, user-defined presets, and the following logs:
•
Keyboard Shadow Log
•
Error Logs
•
Crash Log
•
Vital Product Data
•
DICOM Logs
•
Windows Event Logs
•
Diagnostic Logs
•
Service Logs
•
FibroScan Module Logs
For a detailed list of Service Logs captured, see:
7-3-5 "Capturing Service Logs with ALT+D" on page
.
Figure 7-10 ALT+D Dialog Box
1.)
When Alt+D is pressed, a menu box opens. Enter the following information:
-
System ID serial number.
-
Software version.
-
System Date and time of occurrence.
-
Sequence of events leading to issue.
WARNING
WARNING
(R3.x.x or later) Check “Include Protected Information....” to save all log files.
R1.x.x and R2.x.x
R3.x.x or later
Содержание LOGIQ S8
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4...
Страница 417: ......