Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-3-8-4 Hardware Device Check:
4.) If
X
or
!
is displayed:
-
Left-click on the device with
x
mark, then rigtht-click on
Enable.
-
Left-click on the device with
!
mark, then rigtht-click on
Properties
to perform the device driver
reconfiguration. When some files are required, designate the file in the directories of
C:\WINDOWS\SYSTEM32
or
C:\WINDOWS\SYSTEM32\DRIVERS
. If they are not found,
search them using a windows search function.
5.) Close the window.
6.) Go to
8-3-8-4 Hardware Device Check:
8-3-8-5
LAN Parameter Setting
1.) On the desktop, left-click on
My Network Places
, then right-click on
Properties...
2.) Verify that two network icons appears. (If “Local Area Connection” icon does not appear, the
hardware failure might occur.)
3.) Using the right trackball button, double-click on
Local Area Connection.
4.) When the following window appears, it will be automatically closed later.
Figure 8-137 Network Connections
Figure 8-138 Local Area Connection Enabling