Section 8-16 - Replacement of Key Caps (by special native language keys)
4) Run the OPIO Test tool (ScFpApiTest.exe) in the desktop screen and then click the “Op-panel
Open” button of test tool. (Program path: C:\scanner\target\bin\ScFpApiTest.exe)
Figure 8-104
5) Check on the “Keyboard Test on” check box.
6) Check that pressed key is displayed except for ‘Fn’ key, when AN Keyboard is pressed.
7) End ScFpApiTest by click the “Op-Panel Close” button.
Figure 8-105
8-16-7
Functional Check
Test
Refer to
Note
System Exterior Visual Check
Control panel visual check only
Power On/Off
System Integration Checks
OPIO Test only
OPIO Interface Check